4. Sensor change 

The sensor unit  may need to be changed in service cases. For this purpose the CS 305 has to 
be opened and the sensor cable should be removed at the display terminal first as indicated 
below. Please ensure that the power supply is cut off before opening the casing!

 

 

 

 

The sensor unit can now be screwed out of 
the measuring chamber.  

Remove the green terminal connector from the 
old sensor and connect it to the new sensor. 

 1. Features 

• 

Dew Point Monitor for refrigerant and desiccant dryers (All-In-One) 

• 

Power supply:  

100V…240 VAC / 50…60 Hz (model 0699 3050) 

• 

 

 

  

18 … 32 VDC (model 0699 3051) 

• 

6 mm hose connection for compressed air 

• 

2 relay outputs for alarm 

• 

USB interface for configuration with CSC-350 software 

• 

Wall mountable casing 

• 

Optical alarm on display 

2. Overview 

6 mm hose connector for 
compressed air 

2 cable glands for power 
and alarm/signal cable 

(090) 696-6535 or (093) 820-5599

www.gotek.com.vn

Gotek Vietnam Ltd.

Summary of Contents for CS 305

Page 1: ...n adherence to the instruction manual The warranty is also cancelled once the measuring instrument has been opened provided this is not described in the instruction manual for maintenance purposes Thi...

Page 2: ...pt from liabil ity CS iTEC is not liable for errors which may have slipped into this operating manual CS 305 is designed to meet state of the art safety requirements has been tested and left the facto...

Page 3: ...suring chamber Remove the green terminal connector from the old sensor and connect it to the new sensor 1 Features Dew Point Monitor for refrigerant and desiccant dryers All In One Power supply 100V 2...

Page 4: ...e quired to have a good response time 3 Technical Data Wall mountable casing Size 240 x 160 x 105 5 mm 2 PG inlets cable size 4 5 8 mm Protection class IP 65 Operating temperature 0 50 C Transport tem...

Page 5: ...re quired to have a good response time 3 Technical Data Wall mountable casing Size 240 x 160 x 105 5 mm 2 PG inlets cable size 4 5 8 mm Protection class IP 65 Operating temperature 0 50 C Transport te...

Page 6: ...suring chamber Remove the green terminal connector from the old sensor and connect it to the new sensor 1 Features Dew Point Monitor for refrigerant and desiccant dryers All In One Power supply 100V 2...

Page 7: ...pt from liabil ity CS iTEC is not liable for errors which may have slipped into this operating manual CS 305 is designed to meet state of the art safety requirements has been tested and left the facto...

Page 8: ...n adherence to the instruction manual The warranty is also cancelled once the measuring instrument has been opened provided this is not described in the instruction manual for maintenance purposes Thi...

Reviews: