CS 305
CS 305
CS 305 English, V1.0 10.01.14
CS 305 English, V1.0 10.01.14
0970 0074
0970 0074
5
4
Connect the power supply cable as indi-
cated in the picture. Please refer to the
display user guide for the terminal connec-
tions.
The second cable gland can be used for
alarm and signal cables.
After the wiring is done, the CS 305 can
be closed with 4 screws.
Use the covers to close the screw open-
ings.
Connect the compressed air through a 6
mm hose (we recommend Teflon).
CS 305 will produce a small compressed
air bleeding of about 2 l/min. This is re-
quired to have a good response time.
3. Technical Data
Wall mountable casing
Size: 240 x 160 x 105.5 mm
2 PG inlets, cable size: 4.5 ... 8 mm
Protection class: IP 65
Operating temperature
0 ... 50 ºC
Transport temperature
-20 ... 60 ºC
Weight
520 g
For detailed information about the technical data of the display unit and the dew point sensor
please refer to separate user manuals:
- IM_0073_CS 305 (Display unit)
- IM_0072_CS 217 (dew point sensor)
4. Installation
The instrument is mounted at a wall. Please follow the installation steps as described below.
For the wall mounting please first mount the
brackets to the bottom of the casing as indi-
cated in the picture