background image

 

FA410_V2_01_2008 

4

TECHNICAL DATA 

 
Measuring range 

-80…20 °Ctd pressure dew point resp. dew point in °Ctd 

 

 

 0…100 % RH 

  

-30…70 

°C 

Type 0699.0410, FA 410 

-80…20 °Ctd 

 4…20 mA 

Type 0699.0412, FA 410 

-20…50 °Ctd 

 4…20 mA 

Accuracy:  

typical 

 2 °Ctd of -80…-40 °Ctd 

 

 

 

 1.5 °Ctd of -40…0 °Ctd 

 

 

 

 1 °Ctd of 0…20 °Ctd 

 

 

 

 

 

 

 

 

 

 

 

 

Pressure range: 

-1…50 bar standard 

Power supply: 

16..30 VDC 

 

Output: 

 

4…20 mA 3-wire technology 

Protection class: 

IP 65 

EMV: 

 

DIN EN 61326 

Operating temperature: 

-30…70 °C (ideal 0…50 °C) 

Storage temperature: 

-40…80 °C 

Load for analogue output: 

< 500 Ohm 

Screw-in thread: 

G 1/2" stainless steel 

Material of housing: 

zinc alloy, PC, ABS 

Sensor protection: 

sinter filter 50 

m stainless steel 

Connection: M12, 

5-pole 

Response time t95: 

< 30 seconds (descending) 

 

 

< 10 seconds (ascending) 

 
 

Accuracy FA 410

0,0

0,5

1,0

1,5

2,0

2,5

3,0

-80

-60

-40

-20

0

20

Dew point [°Ctd]

A

ccu

racy

 [

°C

]

Summary of Contents for FA 410

Page 1: ...A 4 41 10 0 Humidity measuring instruments for measurement of the pressure dew point and the atmospheric dew point in different applications compressed air plants refrigerating adsorption dryers gran...

Page 2: ...ustomers buy our high standard products every year There are a few good reasons for doing so Cost performance ratio Reliable quality at a fair price We have the ideal solutions for your measuring task...

Page 3: ...h accuracy of 2 Ctd Calibration on location and testing with CS control and calibration set PC connection set Programming via SFA Software Analogue output 4 20 mA scalable Calibration and adjustment i...

Page 4: ...ard Power supply 16 30 VDC Output 4 20 mA 3 wire technology Protection class IP 65 EMV DIN EN 61326 Operating temperature 30 70 C ideal 0 50 C Storage temperature 40 80 C Load for analogue output 500...

Page 5: ...onnection diagram DIAGRAM OF INSTRUMENT Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 FA 410 Connector plug SDI VB VB I NC Connection cable 0554 0104 5 m 0554 0105 10 m brown white blue black SDI Digital signal inter...

Page 6: ...uring chamber Compressed air flows continuously at 7 bar approx 1 l min expanded in the capillary pipe of the measuring chamber The reaction times for the humidity reading are shorter than in case of...

Page 7: ...stments or similar are carried out the warranty services are free of charge but there is a charge for other services such as transport and packing costs Other claims especially those for damage occurr...

Page 8: ...8052 Villingen Schwenningen Phone 49 0 7705 97 89 9 0 Fax 49 0 7705 97 89 9 20 info cs instruments com www cs instruments com Sales office NORTH Am Oxer 28c D 24955 Harrislee Phone 49 0 461 700 20 25...

Reviews: