FA410_V2_01_2008
4
TECHNICAL DATA
Measuring range
-80…20 °Ctd pressure dew point resp. dew point in °Ctd
0…100 % RH
-30…70
°C
Type 0699.0410, FA 410
-80…20 °Ctd
≙
4…20 mA
Type 0699.0412, FA 410
-20…50 °Ctd
≙
4…20 mA
Accuracy:
typical
2 °Ctd of -80…-40 °Ctd
1.5 °Ctd of -40…0 °Ctd
1 °Ctd of 0…20 °Ctd
Pressure range:
-1…50 bar standard
Power supply:
16..30 VDC
Output:
4…20 mA 3-wire technology
Protection class:
IP 65
EMV:
DIN EN 61326
Operating temperature:
-30…70 °C (ideal 0…50 °C)
Storage temperature:
-40…80 °C
Load for analogue output:
< 500 Ohm
Screw-in thread:
G 1/2" stainless steel
Material of housing:
zinc alloy, PC, ABS
Sensor protection:
sinter filter 50
m stainless steel
Connection: M12,
5-pole
Response time t95:
< 30 seconds (descending)
< 10 seconds (ascending)
Accuracy FA 410
0,0
0,5
1,0
1,5
2,0
2,5
3,0
-80
-60
-40
-20
0
20
Dew point [°Ctd]
A
ccu
racy
[
°C
]