background image

Rear Panel Connections

INPUT POWER CONNECTION

The external power supply provided with the monitor accepts 100 – 
240VAC @ 50 – 60Hz and outputs 12VDC @ 1.0A. This may be plugged
directly into the monitor's power jack. Alternatively, any supply from 7.5 to
48V, AC or DC with a capacity of greater than 10VA may be used.

Power-Over-Ethernet is also supported. An IEEE-802.3AT Power-Over-
Ethernet hub or injector is required. Plug the cable from this device into 
the Ethernet input of the monitor. 

SENSOR CONNECTIONS

Temperature sensors are connected by using the DB9 connectors 
provided. There are two four-wire inputs on each connector. 

Ch

Pin Function

Cable Color

A

8

Excitation C)

White

A

9

Excitation Current(-), 

Green

A

4

Voltage Sense(+)

Red

A

5

Voltage Sense(-)

Black

B

6

Excitation C)

White

B

7

Excitation Current(-)

Green

B

1

Voltage Sense(+)

Red

B

2

Voltage Sense(-)

Black

B

3

Option power, 5VDC

Dual Sensor Cables (Cryo-con part number 4034-038) are provided for 
each input channel. Cable color codes are shown in the table above. 
Note that the shields of this cable are connected to the monitor's chassis 
ground.

IMPORTANT:

 The monitor requires that an Earth 

Ground reference connection is made at the rear 
panel. Failure to provide this connection will result in 
erratic measurements and can even damage input 
circuits.

The sensor cables provided connect their shields to 
the monitor's chassis; Therefore, the required Earth 
Ground can be made by connecting the shield wire at 
the opposite end to a ground reference point. This is 
usually done by connecting it to the back-shell of the 
cryostat connector.

RELAY CONNECTION

Relay connections are made on the rear panel using the 3.5mm, 4-pin 
detachable terminal block provided. 

Pin

Function

1

Relay #1 N.O.

2

Relay #1 Common.

3

Relay #2 N.O.

4

Relay #2 Common.

Terminal block contacts are rated at 10.0A. Relay contact ratings are 
10A@125 VAC, 5A@250VAC or 5A@30VDC

Cryogenic Control Systems, Inc.
PO Box 7012
Rancho Santa Fe, CA 92067

QS18I 10/18

E-mail: [email protected]

Telephone: 858-756-3900

www.cryocon.com