IFU-000100 [Indice C] [SYMS I] [P.
12
/
18
]
Replacing the bands
Carry out your decontamination
procedure.
Recommendation:
Dry the
decontaminated elements before
reassembly.
Reassemble the equipment and
check manually that the cooling jig
comes into contact with the welding
band, otherwise, check the positioning
of the end-stop and adjust the cooling
jig with the help of the knurled nut.
Close the protective flap using the
rapid nut, turning it clockwise.
Open the protective flap using the
rapid nut, turning it anti-clockwise.
Loosen the knurled nut and remove
the cooling jig.
Remove the welding electrode
assembly by pulling it by the levers.