background image

2.2  Rear Panel Input/Output Signals  
Figure 2.1 shows the input and output connectors on the rear panel.

1

2

3

4

6

7

8

9

AC

GND

J8

J2

EXT 10MHZ

IINPUT

RF INPUT

10MHZ REF

OUTPUT

INT

EXT

AUTO

S1

J101

J11

ALARM

J102 - RF OUTPUT
Super SMA (female) 50

Ω

. 20.2 to 21.2 GHz, 

-50 dBm max output signal. 

AC INPUT 
100-240 ±10% VAC, 
47-60 HZ
Uses 2 amp Slow Blo, 
5mm fuse

S1 - 10MHZ SWITCH
3-position switch. INT selects internal 10MHz ref. 
AUTO selects internal 10MHz ref, unless a 
+3dBm ±3dB 10MHz signal is inserted on J2. 
EXT selects external 10MHz ref from J2.

J2 - EXT 10MHZ INPUT
BNC (female) 50

Ω

/75

Ω

. Insert external 

10MHz reference here (3 dBm ± 3 dB).

J101 - RF INPUT
2.92 mm (female) 50

Ω

, 30.0 to 

31.0 GHz, -30.0 to 31.0 GHz to  
dBm max input signal. .

J11 - ALARM
DB9 (female) connector. 
See Table 2.1. 

J8 - 10MHZ REF OUTPUT
BNC (female) 50

Ω

/75

Ω

Outputs the currently selected 
10MHz ref signal (3 dBm ±3 dB).

RF OUT

J102

FIGURE 2.1     2116-31T21 Rear Panel I/O

ʼ

s

TABLE 2.1 J1

11 

TABLE 2.1 J11 Pinouts (DB9)

Pin

Function

1

Not Used

2

Not Used

3

Not Used

4

Not Used

5

GND

6

Alarm Relay: Common

7

Alarm Relay: Normally Open

8

Not Used

9

Alarm Relay: Normally Closed

TABLE 2.2 C

Connector

2.2 

Connector

RF Input

RF Output

STD

2.92 mm

Super SMA, 50

Ω

2.3  Front Panel Indicators 
Figure 2.2 shows the front panel indicators.

DS2 - ALARM LED
Lights RED when PLL comes 
out of lock.

DS1 - POWER LED A
Lights GREEN when DC voltage is 
present from AC power supply.

DS3 - EXT 10MHZ LED
Lights YELLOW when external 
10MHz reference is active.

POWER

EXT

10MHZ ALARM

FIGURE 2.2     2116-31T21 Front Panel Controls and Indicators

2116-31T21 Manual, Rev. 0    

            Page 6

   

    

12/09/19

Summary of Contents for 2116-31T21

Page 1: ...reproduced or duplicated without the prior permission of Cross Technologies Inc for purposes of operating the equipment When ordering parts from Cross Technologies Inc be sure to include the equipment...

Page 2: ...gies Inc s obligation under this warranty is limited to repairing or at Cross Technologies Inc s option replacing parts subassemblies or entire assemblies Cross Technologies Inc shall not be liable fo...

Page 3: ...he AUTO position the internal reference is used unless a 3 dBm 3 dB 10MHz reference signal is connected to the external reference input The 2116 is powered by a 100 240 10 VAC power supply and mounted...

Page 4: ...2 0 dB over frequency band 0 5 dB 40 MHz BW Frequency Sense Non inverting LO Characteristics LO Frequency 9 80 GHz Frequency Accuracy 0 01 ppm max over temp internal reference ext ref input 10 MHz In...

Page 5: ...n be secured to a rack using the 4 holes on the front panel Figure 2 0 shows how the 2116 31T21 is assembled 1 PS IN PIN 1 12 VDC PIN 2 5 VDC PIN 3 5 VDC PIN 4 GROUND PIN 5 GROUND PIN 6 12 VDC PIN 7 2...

Page 6: ...to dBm max input signal J11 ALARM DB9 female connector See Table 2 1 J8 10MHZ REF OUTPUT BNC female 50 75 Outputs the currently selected 10MHz ref signal 3 dBm 3 dB RF OUT J102 FIGURE 2 1 2116 31T21 R...

Page 7: ...d at J2 using the rear panel 3 way switch S1 Figure 2 1 6 If EXT is selected or AUTO is selected and there is a 10MHz 3 dBm signal at J2 check that DS3 yellow Ext 10MHZ is on Figure 2 2 7 Check that a...

Page 8: ...quipment is not compromised Additional space between units may be required E Circuit Overloading Consideration should be given to the connection of the equipment to the supply circuit and the effect t...

Page 9: ...INC 6170 Shiloh Road Alpharetta Georgia 30005 770 886 8005 FAX 770 886 7964 Toll Free 888 900 5588 WEB www crosstechnologies com E MAIL info crosstechnologies com PRINTED IN USA 2116 31T21 Manual Rev...