background image

2.2

  

Rear Panel Input/Output Signals  - 

Figure 2.1 shows the input and output connectors on the rear panel.

    

AC

GND

J8

J2

EXT 10MHZ

RF OUTPUT

INPUT

10MHZ REF

OUTPUT

INT

EXT

AUTO

S1

J1

L-BAND

INPUT

J101

J1 - L-BAND INPUT

BNC (female) 50 

. 0.95 to 1.45 GHz

20 dBm max input signal. See Table 
2.2 for other connector options.

AC INPUT

 

120-240 ±10%VAC, 47-60 HZ
Uses 2 amp Slow Blo, 5mm 
fuse

S1 - 10MHZ SWITCH

3-position switch. INT selects internal 10MHz ref. 
AUTO selects internal 10MHz ref, unless a 0dBm 
±3dB 10MHz signal is inserted on J2. EXT selects 
external 10MHz ref from J2.

J1 - EXT 10MHZ INPUT

BNC (female) 75 

.  Insert external 

10MHz reference here (3 dBm ± 3 
dB).

J101 - RF OUTPUT

Type N (female) 50 

12.2 to 12.7 GHz, -20 to -5 dBm max 
output signal. See Table 2.2 for other 
connector options.

J11 - ALARM

DB9 (female) connector. 
See Table 2.1. 

J8 - 10MHZ REF OUTPUT

BNC (female) 75 

. Outputs currently 

selected 10MHz ref signal (3 dBm ±3 dB).

J11

ALARM

1

2

3

4

6

7

8

9

FIGURE 2.1 2115-122 Rear Panel I/O

ʼ

s

TABLE 2.1 J11 Pinouts (DB9)

TABLE 2.1 J11 Pinouts (DB9)

Pin

Function

1

Not Used

2

Not Used

3

Not Used

4

Not Used

5

GND

6

Alarm Relay: Common

7

Alarm Relay: Normally Open

8

Not Used

9

Alarm Relay: Normally Closed

TABLE  2.2  Connector  Options

TABLE  2.2  Connector  Options

TABLE  2.2  Connector  Options

Option

RF

L-Band

SN

SMA, 50

Type N, 50

SS

SMA, 50

SMA, 50

2.3

  

Front Panel Indicators -

The following are the front panel indicators.

DS2 - ALARM LED

Lights RED when PLL comes 
out of lock.

DS1 - POWER LED A

Lights GREEN when DC 
voltage is present from AC 
power supply.

DS3 - EXT 10MHZ LED

Lights YELLOW when external 
10MHz reference is active.

MODEL 2115

C

ROSS

 T

ECHNOLOGIES INC.

POWER

UPCONVERTER

EXT

10MHZ

ALARM

FIGURE 2.2 2115-122 Front Panel Controls and Indicators

2115-122 Manual, Rev. 0

 

 

Page 6

    

   5/04/09

Summary of Contents for 2115-122

Page 1: ...uced or duplicated without the prior permission of Cross Technologies Inc for purposes of operating the equipment When ordering parts from Cross Technologies Inc be sure to include the equipment model...

Page 2: ...ies Inc s obligation under this warranty is limited to repairing or at Cross Technologies Inc s option replacing parts subassemblies or entire assemblies Cross Technologies Inc shall not be liable for...

Page 3: ...s used in the EXT position the external reference is used and in the AUTO position the internal reference is used unless a 3 dBm 3 dB 10MHz reference signal is connected to the external reference inpu...

Page 4: ...e Non inverting LO Characteristics LO Frequency 11 25 GHz Frequency Accuracy 0 01 ppm max over temp internal reference external reference input 10 MHz Level 3 dBm 3 dB 75ohms External In or Internal o...

Page 5: ...semblies The 2115 122 can be secured to a rack using the 4 holes on the front panel Figure 2 0 shows how the 2115 122 is assembled 1 PS IN PIN 1 12 VDC PIN 2 5 VDC PIN 3 5 VDC PIN 4 GROUND PIN 5 GROUN...

Page 6: ...DB9 female connector See Table 2 1 J8 10MHZ REF OUTPUT BNC female 75 Outputs currently selected 10MHz ref signal 3 dBm 3 dB J11 ALARM 1 5 2 3 4 6 7 8 9 FIGURE 2 1 2115 122 Rear Panel I O s TABLE 2 1...

Page 7: ...at J2 on rear panel switch S1 Figure 2 1 6 If EXT is selected or AUTO is selected and there is a 10MHz 3 dBm signal at J2 check that DS3 yellow Ext 10MHZ is on Figure 2 2 7 Check that a 10MHz 3 dBm 3...

Page 8: ...equipment is not compromised Additional space between unit may be required E Circuit Overloading Consideration should be given to the connection of the equipment to the supply circuit and the effect...

Page 9: ...HNOLOGIES INC 6170 Shiloh Road Alpharetta Georgia 30005 770 886 8005 FAX 770 886 7964 Toll Free 888 900 5588 WEB www crosstechnologies com E MAIL info crosstechnologies com 2115 122 Manual Rev 0 Page...

Reviews: