Incompatible tension level, press      to return.

The door was opened, press     to return.

Canceled by user.

0/100

0,5/111

1,0/120

1,5/127

1,6/128

1,7/129

LED

CAUSE

Heating time limit, press      to return.

Low pressure, press      to return.
Over pressure, press     to return.

10

e

When  the  autoclave  reaches  the  sterilization  temperature/pressure,  it  will  beep  once  and  the  LED 

       

will  blink 

indicating the sterilization for 16 minutes (table 5).

Obs.:

  for  altitudes  higher  than  2000  m,  the  sterilization  will  last  20  minutes.  If  the  autoclave  does  not  reach  the 

proper sterilization temperature in 40 minutes, by lack of water, excess of meterial or air leakage, the cycle will be 
automatically canceled. Voltaje oscilations may cancel the cycle at any time.
If  a  cycle  interruption/cancellation  is  needed,  press        .  In  this  case,  the  LEDs 

Cancelado 

(Canceled)  and 

0,5 

kgf/cm²

 will be on, the autoclave will beep twice and return to the standby mode.

At  the  end  of  the  sterilization,  the  equipment  will  beep  once,  announcing  depressurizing.  The  pressure  LEDs  will 

indicate  0  pressure  at  the  end  of  this  process.  After  that,  the  heating/drying  LED  will  blink  with  2  beeps  every  4 

seconds, indicating the need to slightly open the door, for a proper drying (figures 10 and 11).

ATTENTION!

When opening the autoclave for drying, see the correct 

position of the door (figures 10 and 11). Keep the handle 

all  the  way  down  (fig.  11).  Vitale  Class  performs  drying

 

with  open  door

.  It  is  usual  that  in  this  moment,  some 

steam comes out through the door.

f

ATTENTION!

  Even  after  the  cycle  is  finished,  the  chamber  content  will  be  still  at  high  temperature 

Never 

directly 

touch the chamber, trays or material, wait until the temperature is proper for handling. Remember to use glover for 

security and thermal protection. 

Wait between 15 and 20 minutes for cooling before start a new cycle

.

After drying, the autoclave will beep once and the        LED will keep on. To return to the standby mode, press        . 
If a extra drying is needed, press       to perform this process again.

After use, turn off the equipment pressing          for 2 seconds.

g

h

FAILURE INDICATIONS

Table 6

 

When  a  cycle  is  canceled,  the  Vitale  Class  autoclave  will  light  up  a  LED  related  to  the  possible  cause  of 

canceling, together with the Cancelado (Canceled) LED, according to the table below: 

Fig.10

Fig.11

Upper view

CRISTÓFOLI

A

U

T

O

C

L

A

V

E

S

U

S

E

M

A

N

U

A

L

Summary of Contents for vitale 12

Page 1: ...ll the requirements of the applicable standards promote the continuous improvement of its quality and environmental standards prevent pollution reduce its environmental impact and continuously train its staff achieving therefore a sustainable and profitable maximization of its value Rev 2 BPF Brazilian standard similar to GMP Good Manufacturing Practices FDA US Vitale Class 12 21 Autoclave 13485 D...

Page 2: ...clave Failure Indications Performing a New Initial Cycle Technical Data Quality Control Preventive Maintenance Monitoring Troubleshooting How To Proceed When Service Is Needed Guidance for Final Disposal of the Equipment Warranty Terms Use Instructions Quick Guide 03 04 04 04 05 06 07 08 08 10 11 11 12 12 13 14 14 15 16 Manual Vitale Class Ingl Rev 4 2018 INDEX CRISTÓFOLI A U T O C L A V E S U S E...

Page 3: ...of Manufacture Boas Práticas de Fabricação 4 P B F LOT SN Fragile handle with care Dangerous Electrical Tension Serial Number Lot number Keep out of Sunlight Recyclable Keep Dry ISO 13485 13485 DE S G A E M S E T T Ã S I O S Antimicrobial Painting 135 ºC BPF Brazilian standard similar to GMP Good Manufacturing Practices FDA US CRISTÓFOLI A U T O C L A V E S U S E R M A N U A L ...

Page 4: ...alve opening closing deaerating and depressurizing These processes are part of the proper functioning of the equipment Cristófoli is not responsible for accidents that might occur due to the jolts caused by the noises produced by the autoclave When the handle to open the door the operator must do it easily Never use too much force pulling for it may damage the equipment s internal mechanisms It is...

Page 5: ...NSTRUÇÕES PRESSÃO MÁXIMA DE OPERAÇÃO CAPACIDADE CONFORMIDADE NORMA NBR 11817 Fig 1 5 liters FUSE HOLDER HOW TO IDENTIFY YOUR AUTOCLAVE The Identification label removal will cause automatic loss of warranty STERILIZATION CHAMBER BASE COVER TEMPERATURE SENSOR INTERNAL STEAM EXITS DOOR Panel Lid set LID DOOR GASKET HANDLE TRAY HOLDER TRAYS SAFETY VALVE ANTI VACCUUM PANEL HANDLE KEYBOARD IDENTIFICATIO...

Page 6: ...m table below Table 1 AUTOCLAVE CIRCUIT BREAKER VOLTAGE Vitale Class 12 Vitale Class 21 WIRING GAUGE Fordistancesupto5m 16ft fromthecircuitbreakertotheoutletusedto connecttheequipment use2 5mmwiring From5to15m 16to46ft use4mm andfrom15to50m 46to164ft use6mm NOTE Forthe21 litersmodel 127V fordistancesupto15to35m 49to114ft use6mm wiring and 8mmfor114to164ftdistance 8 A 127 220 V 12 A NOMINAL CURRENT...

Page 7: ... Vitale Class 21 Close the door and press to begin 2 1 The performance of the autoclave is directly related to the atmospheric pressure which varies according to the altitude of where is it installed The following instructions will guide you on how to perform the Initial Cycle of your autoclave This reinforces that your autoclave will efficiently work at the location where it is installed Fig 6 If...

Page 8: ...used fuse is the glass 6 35x32 AGLF fast acting If substitution is needed see Troubleshooting page 13 DOOR KEY Internal device that prevents the functioning of the equipment if the door is not properly closed TEMPERATURE VS PRESSURE DATA CROSSING ELECTRONIC SYSTEM Internal system that automatically cancels the cycle if any problem is detected with pressure reading DOOR LOCKING Internal device whic...

Page 9: ... not exceed 75 of the chamber capacity Place the envelopes always with the paper table 4 side upturned it the steam resulting on a fast and efficient drying and for th s eases penetration process e ame reason boxes and trays must be totally perforated so that allows steam circulation and makes drying easier c d e Press to start the autoclave will beep once and the LED Aquecendo Secando Heating Dry...

Page 10: ...ter that the heating drying LED will blink with 2 beeps every 4 seconds indicating the need to slightly open the door for a proper drying figures 10 and 11 ATTENTION When opening the autoclave for drying see the correct position of the door figures 10 and 11 Keep the handle all the way down fig 11 Vitale Class performs drying with open door It is usual that in this moment some steam comes out thro...

Page 11: ...ze 28 5 mm approximate value 127 220V AC 95 254 V 50 60 Hz 1600 Watts 485 Watts hour 2 216 kPa 2 2 kgf cm 100ºC 15ºC a 40ºC Up to 3500 m 10363350017 12 liters 17 7 kg 19 4 kg Stainless Steel 22 x 30 cm 35 x 34 5 x 48 5 cm 15 5 x 1 2 x 28 cm small 20 x 1 2 x 28 cm large 67 mm approximate value 127 220V AC 95 254 V 50 60 Hz 1200 Watts 336 Watts hour 2 216 kPa 2 2 kgf cm 100ºC 15ºC a 40ºC Up to 3500 ...

Page 12: ...nitorização Física e Química on www youtube com CC available Door gasket cleaning Remove the gasket by pulling it after that clean it with a cloth which does not shed moistened with water and place it on the lid again Sterilizing monitoring biological test Peform biological test every 7 days To obtain more information about how to perform the chemical tests consult www cristofoli com at the sessio...

Page 13: ...ed to the outlet The key was not pressed to turn the autoclave on Blown fuse Verify if there s a power failure in your area Connect the power cord to the outlet properly Press Replace the fuse that is located near to the power chord unplug the autoclave remove the cover of the fuse holder take out the damaged fuse put the new one replace the cover and turn on the equipment THE AUTOCLAVE PRESSURE G...

Page 14: ...possible to determine how long is it as it varies according to the intensity and how the equipment is used or handled Exception made for the component lid which must be replaced every 10 years CRISTÓFOLI EQUIPAMENTOS DE BIOSSEGURANÇA LTDA makes clear its concern already demonstrated by the implementation of the Environmental Management System according to standard NBR ISO 14001 2004 and strongly r...

Page 15: ...y the buyer owner The warranty will be voided in cases of Problems arising from natural causes such as floods lightning etc Use of non distilled water for the operation of the autoclave Damage caused by accidents such as dents drops short circuits fire etc Damage caused by humidity excessive exposure to sun light and salinity Sinister theft or robbery Improper use installation of the equipment and...

Page 16: ... the end of the process the LED will be on Wait until the temperature of the material decreases to handle it 16 USE INSTRUCTIONS QUICK GUIDE Read carefully the Instruction Manual TIEMPO DE SECADO For altitudes above 2 000 m the sterilization time is 20 minutes Maximum drying temperature 129 C Default Program Extra Drying PROGRAMS STERILIZATION TEMPERATURE STERILIZATION PRESSURE STERILIZATION TIME ...

Reviews: