6.39. Before starting an assessment
Turn the instrument on by pressing button 3. Before starting use, it is important to
check and adjust the sensitivity value of the sensor used. To access sensitivity settings,
press button 6 for 3 seconds, select the option System with button 6, and then select
the option Sensor by using button 6. Change the sensor type by using buttons 4 and 5
to alternate between the options and confirm it by pressing button 6. Select the option
Sensitivity by using buttons 4 and 5 and confirm it by pressing button 6. Press buttons
4 and 5 to increase or decrease the sensitivity value. To advance to the next parameter,
press button 6. Repeat the procedure for all axes.