Maintenance
▶ Cleaning and Disinfection
Contaminated CPR BAND
(Model: CREDO-CB)
should be properly
cleaned and disinfected.
Cleaning
1. Prepare a small brush (e.g. toothbrush) moistened with warm water
(40 - 50 °C).
2. Scrub the device with a wet brush for at least 2 minutes. Do not
clean the charging port directly.
3. If the port is clogged, use the brush to remove any obstruction.
4. Wipe the exterior with a soft cloth dampened in lukewarm water
(22 - 40°C).
Disinfection
1. The exterior can be disinfected in a 0.55% solution of ortho-
phthalaldehyde, or a 70% solution of isopropyl alcohol
(isopropanol). Spray the exterior surfaces with the solution and
ensure it remains wet for a minimum of 12 minutes. Repeat
spraying as necessary to account for evaporation. Do not spray the
solution directly on the charging port.
2. Wipe the exterior at least 3 times with a clean damp cloth to
remove all traces of the disinfectant.
3. Allow to dry completely.
Warning
Do not immerse the CPR BAND
(Model: CREDO-CB)
in running water or
allow moisture to penetrate it. Do not sterilize the CPR BAND