30
Maintenance
Storage after use
▶ Routine maintenance
If the button is pressed while the charger is connected, CPR BAND
(Model: CREDO-CB)
is rebooted and all setting values are initialized.
The frequency of maintenance depends on usage, if it is used daily, it
is once after 4300 hours.
▶ After each use
After use on a patient, the CPR BAND
(Model: CREDO-CB)
may be
contaminated and should be handled appropriately.
1. Place the contaminated CPR BAND
(Model: CREDO-CB)
in a plastic
bag until it can be cleaned.
2. If it is visibly soiled, wipe the CPR BAND
(Model: CREDO-CB)
with a
soft cloth or paper towel to remove as much contamination as
possible.
3. Separate the module and the band. The upper part of the module
must be separated first.
4. Clean the CPR BAND
(Model: CREDO-CB)
as described under
Cleaning and Disinfection. Proper cleaning is required to achieve
disinfection.
5. Check the exterior of the CPR BAND
(Model: CREDO-CB)
for signs of
damage. Contact CREDO for replacement parts or service if
needed.