3C
Interfacing and Programming
CPI Canada Inc.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-42
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
3C.7.3 Receptor Setup
The
RECEPTOR SETUP
function allows each of the image receptors to be programmed as defined in the table following the example menu
screens.
The
RECEPTOR SETUP
menus for the membrane console are shown below.
MENU 1
* RECEPTOR SETUP [sym] *
TUBE: 1
TOMO BUT: 2500 MS
RECEPTOR SYM: [sym]
INTERFACE OPTS: 0
MEMORY: DEF
+
TOMO: NO
-
EXIT >>
Note: The
INTERFACE OPTS
function is available only on CMP200
®
DR generators.
MENU 2
* RECEPTOR SETUP [sym]*
SF/LF SWITCH: MAN
DEFAULTS
AEC BACKUP: FIXED
AEC CHANNEL: 1
AEC BACKUP MAS: 100
+
AEC BACKUP MS: 1000
-
<< >>