Enclosure Replacement
12-32
Service Manual
34.
Place a cable tie around the fan cable and thermistor cable approximately 10 cm
(4 inches) from the Power PCBA connectors (
). Trim the end of the
cable tie.
Figure 12-29.
Cable Ties and Routing of Cables
35.
Make sure that thermostat cable is routed as shown in
. It must be
positioned to the inside of the adjacent screw boss and pole clamp mounting bolt
and inside the tabs around the fan assembly.
36.
Using two cable ties, secure the thermostat cable to the heater cable as follows
(
):
a.
Place one cable tie around the cables approximately 5 cm (2 inches) from
where the heater cable exits below the fan assembly.
b.
Place the other cable tie around the cables approximately 10 cm (4 inches)
from the heater cable connector (P114) on the Power PCBA.
c.
Trim the ends of the cable ties.
1
Cable Ties
(Thermostat Cable to Heater Cable)
3
Thermostat Cable
Routing Tabs
2
Cable Tie
(Fan Cable to Thermistor Cable)
Summary of Contents for WarmTouch
Page 1: ...Service Manual WarmTouch Convective Warming Unit...
Page 8: ...Page Left Intentionally Blank vi Service Manual...
Page 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Page 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Page 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Page 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Page 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Page 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Page 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Page 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Page 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Page 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Page 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Page 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Page 313: ......