Internal Component Replacement (Inside Rear Enclosure)
11-54
Service Manual
outside of the enclosure. Tighten the screws to 0.18 to 0.27 N-m (1.6 to 2.4
lb-in).
24.
Remove the strap or cable tie from the hose duct adapter.
25.
Reconnect the cables to the Power PCBA as follows (
a.
Connect the heater cable to the 5-pin P114 connector. Make sure that the
connector is oriented as shown in
, and that it is fully seated.
b.
Connect the thermostat cable to the 2-pin P113 connector. Make sure that
the connector is oriented as shown in
, and that it is fully seated.
c.
Connect the fan cable to the 4-pin P111 connector. Make sure the connector
latch engages.
d.
Connect the thermistor cable to the 2-pin P108 connector. Make sure the
connector latch engages.
26.
Make sure that thermostat cable is routed as shown in
. It must be
positioned to the inside of the adjacent screw boss and pole clamp mounting bolt
and inside the tabs around the fan assembly.
Figure 11-52.
Routing of Cables around Fan Assembly
1
Cable Ties
(Heater Cable to Thermostat Cable)
2
Thermostat Cable
Routing Tabs
Summary of Contents for WarmTouch
Page 1: ...Service Manual WarmTouch Convective Warming Unit...
Page 8: ...Page Left Intentionally Blank vi Service Manual...
Page 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Page 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Page 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Page 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Page 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Page 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Page 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Page 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Page 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Page 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Page 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Page 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Page 313: ......