Pole Clamp and Mounting Bolt Replacement
Service Manual
11-19
To replace a pole clamp and mounting bolt:
1.
At the back of the unit, completely loosen the knob on the clamp to be replaced.
Remove the knob and clamp foot from the mounting bolt (
Figure 11-17.
Pole Clamp Foot and Knob
2.
Follow the procedure
Separating the Front and Rear Enclosures
3.
Remove the pole clamp mounting bolt as follows (
):
a.
Position the rear enclosure so that you can access the end of the mounting
bolt.
b.
Use a small rubber mallet or similar tool to drive the mounting bolt toward the
inside of the enclosure until you can grasp it and remove it.
1
Clamp Foot (x3)
2
Clamp Knob (x3)
Summary of Contents for WarmTouch
Page 1: ...Service Manual WarmTouch Convective Warming Unit...
Page 8: ...Page Left Intentionally Blank vi Service Manual...
Page 32: ...Page Left Intentionally Blank 1 18 Service Manual Introduction...
Page 42: ...Installation 3 4 Service Manual Figure 3 3 Power Cord Wrapped and Secured...
Page 45: ...Installing the Warming Unit Service Manual 3 7 Figure 3 5 Warming Unit Mounted on IV Pole...
Page 49: ...Installing the Warming Unit Service Manual 3 11 Figure 3 9 Warming Unit Mounted on Cart...
Page 50: ...Page Left Intentionally Blank 3 12 Service Manual Installation...
Page 70: ...Page Left Intentionally Blank 5 12 Service Manual Maintenance...
Page 84: ...Page Left Intentionally Blank 6 14 Service Manual Troubleshooting...
Page 105: ...Power Cord Replacement Service Manual 8 7 Figure 8 5 Power Cord Wrapped and Secured...
Page 118: ...Page Left Intentionally Blank 9 8 Service Manual Accessing the Inside of the Warming Unit...
Page 246: ...Enclosure Replacement 12 14 Service Manual Figure 12 10 Power Cord Wrapped and Secured...
Page 266: ...Page Left Intentionally Blank 12 34 Service Manual Enclosure Replacement...
Page 274: ...Page Left Intentionally Blank 13 8 Service Manual Product Specifications...
Page 313: ......