Service and repair
8-24
10070389 Rev. A
Puritan Bennett™ 700 Series Ventilator System Service Manual
Figure 8-19. Disassembling Communications option assembly
(remote alarm, nebulizer, nurse call, and serial port hardware)
8.6.3.2 Nebulizer harness and connector
(See Figure 8-19 and Figure 8-20). Remove the Communications option nebulizer harness as
follows. Install by reversing removal procedure.
1.
Remove the Communications option assembly and disconnect the ribbon cable
(Section 8.6.3).
2.
Remove the remote alarm harness and connector (Section 8.6.3.1).
3.
Disconnect the nebulizer harness from the J3 connector on the PCB assembly.
4.
Use a 14-mm open-end wrench to loosen the lock nut that holds the nebulizer connector
to the option plate.
5.
Push the harness and connector through the plate to remove.
8.6.3.3 Nurse call (central station) harness and connector
(See Figure 8-19 and Figure 8-20). Remove the Communications option nurse call harness as
follows. Install by following the instructions that come with the replacement nurse call
harness.
1.
Remove the Communications option assembly and disconnect the ribbon cable
(Section 8.6.3).
2.
Disconnect the nurse call harness from the J4 connector on the PCB assembly.
3.
Use a cutting tool to cut the harness’ three-pin connector from the harness (so you can
pull the harness through the clip inside the option panel).
4.
Use a 10-mm open-end wrench to loosen the lock nut that holds the nurse call connector
to the option plate.
5.
Push the harness through the plate to remove.
7-00254
Remote alarm
harness/connector
Nurse call
harness/connector
PCB assembly
Ribbon cable
Nebulizer
harness/connector
Serial port
hardware