![Covidien Puritan Bennett 700 Series Service Manual Download Page 281](http://html1.mh-extra.com/html/covidien/puritan-bennett-700-series/puritan-bennett-700-series_service-manual_2673918281.webp)
Service and repair
Puritan Bennett™ 700 Series Ventilator System Service Manual
10070389 Rev. A
8-23
Figure 8-18. Removing Communications option assembly
8.6.3.1 Remote alarm harness and connector
(See Figure 8-19 and Figure 8-20). Remove the remote alarm harness as follows. Install by
reversing removal procedure.
1.
Remove the Communications option assembly and disconnect the ribbon cable
(Section 8.6.3).
2.
Disconnect the remote alarm harness from the J2 connector on the PCB assembly.
3.
Use a 19-mm open-end wrench to loosen the lock nut that holds the remote alarm
connector to the option plate.
4.
Push the harness and connector through the plate to remove.
7-00011
Summary of Contents for Puritan Bennett 700 Series
Page 1: ...Service Manual Puritan Bennett 700 Series Ventilator Systems TM...
Page 439: ......