Maintenance, Overhaul, and Software Upgrade
2-14
Service Manual
2.7.8
Inhalation Outlet Check Valve
Figure 2-11.
Inhalation Outlet Check Valve
To replace the inhalation outlet check valve, refer to
and perform the following steps:
1.
Remove the mixing block by removing the two retaining screws (see
2.
Remove the inhalation adapter from the inhalation outlet assembly (see
).
3.
Using your finger, push out the inhalation outlet check valve (see
4.
Replace the inhalation outlet check valve diaphragm (VLV2100M).
5.
Reinstall the components and assembly by reversing steps 1-3.
2.8
Overhaul Procedures
WARNING:
Disconnect electrical power, air and oxygen sources before attempting any disassembly.
2.8.1
Inlet Water Trap Filter and O-Ring Assembly
To replace the inlet filter and o-ring assemblies, follow the procedure in section
Trap Filters and O-Ring Assembly
1
Inhalation outlet check valve
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......