Maintenance Procedures
Service Manual
2-11
Note:
If your ventilator does not have the in-line disk filters already installed, contact the Technical Service Department
to obtain the necessary filter adapters.
2.7.6
Coin Battery (SBC2100P and SBC2108P)
To replace the coin battery, perform the following steps:
1.
For first-time installation only: Locate JP1 on the SBC board and move the jumper from pins 2 and 3 to
pins 1 and 2 (
).
Figure 2-7.
JP1 at Pins 1 and 2
2.
Remove the cable clamp securing the SBC battery (CBL2132A).
3.
Remove the SBC battery from J1 on the SBC board.
4.
Reinstall the battery by reversing steps 2 and 3. When reconnecting the SBC battery to J1, confirm the
cable is connected as shown in
Figure 2-8.
J1 with CBL2132A Connected
5.
Reset the date and time on the ventilator:
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......