Product Views
Operator’s Manual
2-5
6
Alarm silenced icon
Audio OFF icon
The yellow icon indicates Alarm Silenced. This indicator also
shows the time remaining in the alarm silence period.
The red icon indicates Audio OFF.
7
Pulse rate real-time value
Displays the pulse rate in beats per minute. Current upper and
lower alarm limit settings appear as smaller values to the left of
the dynamic pulse rate value.
8
Battery status icon
Displays the battery charge remaining on the internal 5-hour or
optional 10-hour battery.
•
Charged Battery —
A steady green battery icon indicates
the monitoring system is running on internal battery
power and the battery is fully charged.
•
Low Battery —
A low priority alarm occurs when the
remaining battery power is only enough for 15 minutes of
operation. The flashing yellow alarm message
Low
Battery
appears. Users cannot silence this alarm while
running on battery power. Connect the monitoring
system to AC power to stop the alarm.
•
Critically Low-Battery —
A high priority alarm occurs for
about five (5) minutes before the monitoring system shuts
off. The flashing red alarm message
Critically Low-
Battery
appears. When no charge remains, the moni-
toring system automatically shuts down. Connect the
monitoring system to AC power to avoid any loss of trend
data or settings.
9
AC power indicator
Lights continuously when connected to AC power.
10
Battery charge indicator
Lights when the monitoring system is charging the internal 5-
hour or optional 10-hour battery.
11
Interference indicator
Lights when the monitoring system detects degraded quality in
the incoming signal. Lights intermittently as the monitoring
system dynamically adjusts the amount of data required for
measuring SpO
2
and pulse rate. Lights continuously when the
monitoring system has extended the amount of data required
for measuring SpO
2
and pulse rate. In this case, fidelity in track-
ing rapid changes in these values may be reduced.
1
12
Sensor off indicator
Appears when the sensor is not on the patient.
13
Sensor disconnect indicator
Appears when the sensor is not connected to the monitoring
system.
Summary of Contents for Nellcor Series
Page 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank x...
Page 20: ...Introduction Page Left Intentionally Blank 1 8 Operator s Manual...
Page 36: ...Installation Page Left Intentionally Blank 3 8 Operator s Manual...
Page 118: ...Product Specifications Page Left Intentionally Blank 11 18 Operator s Manual...
Page 156: ...Clinical Studies Page Left Intentionally Blank A 38 Operator s Manual...
Page 159: ......