Product Specifications
11-16
Operator’s Manual
Table 11-13.
Patient Leakage Current
Patient Leakage Current
Condition
AC Line Polarity
Neutral Line
Power Line Ground
Cable
IEC 60601-1
ANSI/AAMI
ES 60601-1
Normal
Normal
Closed
Closed
100 µA
Single Fault
Open
Closed
500 µA
Closed
Open
Normal
Reversed
Closed
Closed
100 µA
Single Fault
Open
Closed
500 µA
Closed
Open
Patient Leakage Current - Mains on Applied Part
Condition
AC Line Polarity
Neutral Line
Power Line Ground
Cable
IEC 60601-1
ANSI/AAMI
ES 60601-1
Single Fault
Normal
Closed
Closed
5000 µA
Reversed
Closed
Closed
Summary of Contents for Nellcor Series