Manufacturer’s Declaration and Guidance
Operator’s Manual
11-15
Table 11-12.
Earth Leakage and Touch Current
Earth Leakage Current
Condition
AC Line Polarity
Line Cord
Neutral
Line Cord
IEC 60601-1
ANSI/AAMI ES 60601-1
Normal
Normal
Closed
Closed
5 mA
Single Fault
Open
Closed
10 mA
Closed
Open
Normal
Reversed
Closed
Closed
5 mA
Single Fault
Open
Closed
10 mA
Closed
Open
Touch Current
Condition
AC Line Polarity
Neutral
Line Cord
Power Line
Ground
IEC 60601-1
ANSI/AAMI ES 60601-1
Normal
Normal
Closed
Closed
100 µA
Single Fault
Open
Closed
500 µA
Closed
Open
Normal
Reversed
Closed
Closed
100 µA
Single Fault
Open
Closed
500 µA
Closed
Open