Installation
5-4
Service Manual
Caution:
Use only a hospital-grade power cord.
Caution:
Ensure the monitoring system is properly grounded when operating on AC
power. If uncertain whether the AC outlet is properly grounded, disconnect
the monitoring system from the outlet and use battery power. Contact a
qualified electrician to examine the outlet for ground connections.
Caution:
Do not block cooling vents.
Ensure the monitoring system remains connected to an AC power source
when not in use so a fully charged battery remains available for use at any time.
To connect to an AC power source
1.
Plug the female connector end of the power cord into the power connector on
the rear of the monitoring system. Reference
2.
Plug the male connector of the power cord into a properly grounded AC outlet.
3.
Verify the monitoring system’s AC power indicator lights.
Note:
If the AC power indicator does not light, check the power cord, user-accessible fuses,
and AC power outlet.
5.3.3
Battery Insertion
WARNING:
Install only Covidien-approved batteries.
The monitoring system ships with a separate internal battery. The battery must
be installed prior to use in a clinical setting. Reference
, p. 12-7. Insert the battery and test the monitoring system
prior to use in a clinical setting. Users should immediately and completely
charge the battery prior to clinical use or temporary storage of the battery.
Users should also remain vigilant when running on battery power and recon-
nect to AC power during a low battery state.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......