- 38 -
8. Warranty
New Cosmos Electric Company Limited (New Cosmos) offers the following as the sole
and exclusive limited warranty available to Customer.
This warranty is in lieu of, and customer waives, all other warranties of any kind or
nature, expressed or implied, including without limitation, any warranty for
merchantability or fitness for a particular purpose. The remedies set forth herein are
exclusive.
New Cosmos warrants to the original purchaser and no other person or entity
(customer) that gas detection product supplied by New Cosmos shall be free from
defects in materials and workmanship for a period of one (1) year from the date of
purchase. This warranty does not include consumables, such as fuses, filters, etc.
Certain other accessories not specifically listed here may have different warranty
periods.
After examination of allegedly defective product return to New Cosmos, with freight
prepaid, should the product fail to conform to this warranty, customer’s only remedy
and New Cosmos’s only obligation shall be, at New Cosmos’s sole option, replacement
or repair of such non-conforming product or refund of the original purchase price of the
non-conforming product. In no event will New Cosmos be liable for any other special,
incidental or consequential damages or losses of any kind whatsoever, including but
not limited to, loss of anticipated profits and any other loss caused by reason of
non-operation of the product.
This warranty is valid only if the product is maintained and used in accordance with
New Cosmos’s instructions and /or recommendations. New Cosmos shall be released
from all obligations under this warranty in the event repairs or modifications are made
by persons other than its own or authorized service personnel or if the warranty claim
results from physical abuse or misuse of the product.
- 37 -
MENU
7. Troubleshooting
Before requesting repair, please refer to the table below.
* If the detector fails to operate, remove the battery and install it again, then turn on the
detector.
Problem Cause
Steps Reference
Pressing the
[ POWER]
button does not
turn on the power
Battery orientation
incorrect
Remove the battery
and reinstall it in a
correct orientation.
“Battery
Replacement” on
page 28
“Charge battery”
on page 29
Battery is depleted
Replace/charge the
battery.
Audio alarm does
not sound
Audio is set to the OFF
position.
Set the audio to the
ON position.
“Audio on/off” on
page 20
Error message
displayed
“Error Messages” on page 27
“0” or “0.0” is
flashing for gas
concentration
The reading shifts in the
negative area. It
indicates there is a
possibility that air
adjustment was
performed in a gas
atmosphere or the unit
was exposed to high
gas concentration gas.
Perform air
adjustment in clean
air.
“Air Adjustment
(Zeroing)” on
page 14