Contact your local Cosmo representative for scheduling Annual Calibration Service.
The following items will be inspected and calibrated.
1) Check the oil mist separator and the filter.
2) Leak check of the leak tester
3) Check the DPS offset.
4) Check the DPS span
5) Check the PS offset.
6) Check the PS span.
4
4
F
F
e
e
a
a
t
t
u
u
r
r
e
e
s
s
f
f
o
o
r
r
M
M
a
a
i
i
n
n
t
t
e
e
n
n
a
a
n
n
c
c
e
e
4.1 K(Ve) Check
Compares K(Ve) measured with a reference master tested part to the K(Ve) stored in memory.
This can be used for daily sensitivity checks of Type J models.
K(Ve) Check behavior will be same as the K(Ve) Automatic Setup.
K(Ve) Check Limit
Setting a tolerance in percentage (±) to the K(Ve) currently stored in memory.
> Main Menu > K(Ve) > K(Ve) Settings > Basic > K(Ve) Check Limit
Manual Operation
1) > Main Menu > K(Ve) > K(Ve) Check
2)
Press
►
to start K(Ve) Check.
Remote Operation
Transmit K(Ve) Check signal and Start Signals through control I/O port.
Results for K(Ve) Check
Lower than LL
Within the Limits
Larger than UL
DET LL
Pass
DET UL