¿ -J1
-
Option -J1 units,the Input and Output connector is VH
connectors(Mfr:J.S.T).
¡
LMA100F
I/O Connector
Mating connector
Terminal
CN1 B3P5-VH
VHR-5N
Chain SVH-21T-P1.1
Loose BVH-21T-P1.1
CN2
B8P-VH
VHR-8N
Chain SVH-21T-P1.1
Loose BVH-21T-P1.1
(Mfr:J.S.T)
¡
LMA150F,LMA240F
I/O Connector
Mating connector
Terminal
CN1 B3P5-VH
VHR-5N
Chain SVH-21T-P1.1
Loose BVH-21T-P1.1
CN2
B6P-VH
VHR-6N
Chain SVH-21T-P1.1
Loose BVH-21T-P1.1
CN3
B7P-VH
VHR-7N
Chain SVH-21T-P1.1
Loose BVH-21T-P1.1
(Mfr:J.S.T)
¿
-S
-
-SN
-
-S indicates a type with chassis, and -SN indicates a type with
chassis and cover (Refer to external view).
Please contact us about the detail of derating curve.
¿
-R
-
You can control output ON/OFF remotely in Option -R models.
To do so, connect an external DC power supply and apply a
voltage to a remote ON/OFF connector, which is available as
option.
Model Name
Built-in
Resistor
Ri [
W
]
Voltage between RC (+)
and RC (-) [V]
Input
Current
[mA]
Output ON Output OFF
LMA100F, LMA150F
LMA240F
780
4.5 - 12.5
0 - 0.5
20max
Ri
RC (-)
RC (+)
2
1
*
1
R
SW
External Power
Source
Remote ON/OFF connector (Optional)
Input Current
Inside of a Power
Supply
Fig.6.2 Example of using a remote ON/OFF circuit
-
Dedicated harnesses are available for your purchase. Please
see Optional Parts for details.
*
1 If the output of an external power supply is within the range
of 4.5 - 12.5V, you do not need a current limiting resistor R. If
the output exceeds 12.5V, however, please connect the cur-
rent limiting resistor R.
To calculate a current limiting resistance value, please use the fol-
lowing equation.
R[
W
]=
Vcc-(1.1+Ri
X
0.005)
0.005
*
Please wire carefully. If you wire wrongly, the internal components
of a unit may be damaged.
¡
Remote ON/OFF circuits (RC+ and RC-) are isolated from input,
output and FG.
¿
-R2
-
The usege is the same as option -R, please refer to Option -R.
-
Reducing standby power is possible by OFF signal of the re-
mote control.
-
Start up time by ON signal in remote control is 350ms(typ).
-
The latch condition in overvoltage protection is removed by tog-
gling ON/OFF signal of remote control.
-
Standby power
LMA100F, LMA150F, LMA240F
0.2Wtyp (AC100V), 0.7Wtyp (AC200V)
¿
-P
-
Constant current mode for battery charging is possible by set-
ting the over current activation point within rated output current.
-
Over current activation point varies depending on output voltage
setting as shown in Table 6.2.
-
Parallel operation is not possible.
Table 6.2 Overcurrent protection
*
Model
Output voltage
setting [V]
Overcurrent
protection [A]
LMA100F-24-PY
19.2
[
Vo<21.6
2.5 - 4.2
21.6
[
Vo<26.4
2.2 - 4.0
26.4
[
Vo<27.5
2.1 - 3.8
LMA150F-24-PY
19.2
[
Vo<21.6
4.3 - 6.2
21.6
[
Vo<26.4
4.1 - 6.1
26.4
[
Vo<27.5
4.0 - 5.9
LMA240F-24-PY
19.2
[
Vo<21.6
7.2 - 9.9
21.6
[
Vo<26.4
6.9 - 9.7
26.4
[
Vo<27.5
6.7 - 9.5
*
-P option can not generate rated output current as shown on label.
6.2 Others
¡
This power supply is the rugged PCB type. Do not drop conduc-
tive objects in the power supply.
¡
At light load, there remains high voltage inside the power supply
for a few minutes after power OFF.
So, at maintenance, take care about electric shock.
¡
This power supply is manufactured by SMD technology. The
stress to PCB like twisting or bending causes the defect of the
unit, so handle the unit with care.
-
Tighten all the screws in the screw hole.
-
Install it so that PCB may become parallel to the clamp face.
-
Avoid the impact such as drops.
¡
While turning on the electricity, and for a while after turning off,
please don’t touch the inside of a power supply because there are
some hot parts in that.
¡
When a mass capacitor is connected with the output terminal (load
side), the output might become the stop or an unstable operation.
Please contact us for details when you connect the capacitor.
AC-DC Power Supplies Medical Type
Instruction Manual
LMA-16
June 29, 2020