11
NUV
ANT
™
MCT Instructions for Use
ENGLISH
Specifications, Compliance and Symbols
Specifications
The following performance specifications are at 20°C (68°F) unless otherwise stated
PiiX
zLink
Patient
Trigger
Magnet
Shelf life
Refer to PiiX
pouch label
N/A
N/A
Battery Charger
Power Requirement
N/A
100-240VAC,
50/60Hz
N/A
Battery Life
7.5 days (180
hrs), non-
rechargeable
Provides 12 hrs of
function before
recharging
N/A
Operating
Temperature
0˚C to 41˚C
(32˚F to 105.8˚F)
0˚C to 45˚C
(32˚F to 113˚F)
N/A
Maximum
Temperature of the
Applied Part
44˚C (111.2˚F)
N/A
N/A
Storage Temperature
(power off)
-20˚C to 65˚C
(-4˚F to 149˚F)
-20˚C to 65˚C
(-4˚F to 149˚F)
N/A
Operating Humidity
10% to 95%
10% to 95%
N/A
Storage Humidity
5% to 95%
5% to 95%
N/A
ECG
Sampling Rate
Digital Resolution
Input Dynamic Range
Input Offset
Dynamic Range
200Hz (+/-5%)
10bits
+/- 5mV
+/- 300mV
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
Impedance
Measurements
Peak current injection
RMS current injection
40 uA
29 uA
N/A
N/A
N/A
N/A
Measurement
Ranges, Heart Rate
25 to 250 BPM
N/A
N/A
Data Storage
Capacity
7.5 days
N/A
N/A
Weight
50g / 1.8oz max
150g / 5.3oz max
Portable
Communication
Means
Bluetooth
between PiiX and
zLink
Cellular Phone
between zLink and
Server
N/A
Arrhythmia Detection Algorithms and Automatic ECG Collection
In addition to patient-driven collection of ECGs using the Patient Trigger Mag-
net, the NUVANT MCT system also uses proprietary algorithms based on rate,
rhythm and morphology to continuously analyze rhythm abnormalities and
to initiate automatic ECG collection. ECGs are automatically collected upon
detection of the following conditions:
•
Heart Rate >= 130 bpm
•
Heart Rate <= 40 bpm
•
Pause >= 3 seconds
•
Atrial Fibrillation
•
Ventricular Tachycardia/Ventricular Fibrillation