![Corsano CardioWatch 287-2 Instruction Manual Download Page 37](http://html1.mh-extra.com/html/corsano/cardiowatch-287-2/cardiowatch-287-2_instruction-manual_2669687037.webp)
As of 2022-11-17
Ι Revision status: 12
37
Guidance and manufacturer’s declaration—electromagnetic
immunity
Corsano racelet is intended for use in the electromagnetic environment specified below. The
customer or the user of Corsano bracelet should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
Discharge
(ESD) IEC
61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast
transient/burs
t IEC 61000-
4-5
±2 kV for
power
supply
lines
±1 kV for
input/outpu
t lines
±2 kV for power
supply lines
±1 kV for
input/output lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge IEC
61000-4-6
±1 kV
differential
mode
±2 kV
common
mode
±1 kV
differential
mode
±2 kV common
mode
Mains power quality should be that
of a typical commercial or hospital
environment.
Power
frequency
(50/60Hz)
magnetic
field IEC
61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.