33
professional upon request.
The pulse oximeter measures the pulse rate signal and
computes a quality factor. When it is not able to perform a good
measurement, it cannot display a heart rate value and displays a
dotted line on the Heart Rate graph.
The operation of the pulse oximeter can also be verified by using
a simulator device (like the
WhaleTeq HRS200
for example).
1. Place the CardioWatch 287-2B device on the simulator
device
2. Select the simulation heart rate (70 bpm for example)
3. Read the value in the Corsano Trials APP, on the heart
rate grap
h, “LAST BPM”. (The error shall be less than
±1 bpm)