18 – Corning Lambda EliteTouch Pipettors
INSTRUCTION MANUAL – 19
Single-channel pipettors 100-1000 μL (Figure 10C)
• Unscrew the Shaft Cap (M) and slide it off the Seal (P) at
the end of the Plunger (I)
• If necessary, remove the O-ring (O) from the Shaft Cap (M)
remove the Plunger (I) with the Seal (P) and pipetting
Spring (N) from the shaft
• Remove the Shaft (J) from the Connecting Bush (L)
• If necessary, remove the Seal (P) from the Plunger (I)
check the condition of components and clean as
recommended
• Replace any damaged parts
• Assemble the components in reverse order
When inserting the shaft
or the shaft holder set into
the Connecting Bush, make
sure that they are inserted all
the way to the end by gently
rotating the shaft or the shaft
holder while pulling on its end.
Incorrect assembly of the shaft or the shaft holder set
in the threaded bush assembly may result in damage to the
components.
After cleaning the plunger, apply onto its surface a thin layer
of lubricant attached to each pipettor.
NOTE:
After replacing the plunger, it is necessary to
recalibrate the pipettor. In the case of any signs of incorrect
operation of the pipet, check the correctness of assembly.
✓
✕
J
H
P
L
M
N
O
I
10C
Summary of Contents for Lambda EliteTouch
Page 16: ......