18 – Corning Lambda EliteTouch Pipettors 

INSTRUCTION MANUAL – 19

Single-channel pipettors 100-1000 μL (Figure 10C)

•  Unscrew the Shaft Cap (M) and slide it off the Seal (P) at 

the end of the Plunger (I)

•  If necessary, remove the O-ring (O) from the Shaft Cap (M) 

remove the Plunger (I) with the Seal (P) and pipetting 

Spring (N) from the shaft

•  Remove the Shaft (J) from the Connecting Bush (L)
•  If necessary, remove the Seal (P) from the Plunger (I) 

check the condition of components and clean as 

recommended

•  Replace any damaged parts
•  Assemble the components in reverse order
When inserting the shaft 

or the shaft holder set into 

the Connecting Bush, make 

sure that they are inserted all 

the way to the end by gently 

rotating the shaft or the shaft 

holder while pulling on its end.
Incorrect assembly of the shaft or the shaft holder set 

in the threaded bush assembly may result in damage to the 

components.
After cleaning the plunger, apply onto its surface a thin layer 

of lubricant attached to each pipettor.

NOTE: 

After replacing the plunger, it is necessary to 

recalibrate the pipettor. In the case of any signs of incorrect 

operation of the pipet, check the correctness of assembly.

J

H

P

L

M

N

O

I

10C

Summary of Contents for Lambda EliteTouch

Page 1: ...Corning Lambda EliteTouch Pipettors Single channel 8 channel and 12 channel Pipettors Instruction Manual...

Page 2: ...6051 0 5 10 20 6052 2 20 50 6053 5 50 100 6054 10 100 200 6055 20 200 1000 6056 100 1000 8 channel and 12 channel Pipettors 8 10 12 10 6057 6061 0 5 10 8 50 12 50 6058 6062 5 50 8 200 12 200 6059 606...

Page 3: ...ector bar in the multi channel pipettors Materials Used Handle parts Polypropylene PP Polycarbonate PC Polyacetal POM Stainless steel Shaft and piston area parts Polypropylene PP Polyvinylidene fluori...

Page 4: ...e the pipettor as it may damage the pipettor The single use of the tip reduces the risk of contamination of samples Keep the pipettor clean avoiding the use of aggressive substances e g acetone for cl...

Page 5: ...ower position A B C The accuracy and precision are obtained gravimetrically using Corning pipet tips performing at least 10 measurements of distilled water at a temperature of 20 C 1 C according to th...

Page 6: ...unter operating mode switch the calibration switch located under the CAL sign on the counter overlay This switch should only be used when the pipettor is calibrated upper position marked as CAL on the...

Page 7: ...film remains relatively constant in successive pipetting operations with the same tip this error can be eliminated by forming the film before transferring the first sample This is done by aspirating a...

Page 8: ...holding the handle unscrew the remaining part of the shaft holder set from it Then proceed according to the pipettor models shown below See ISO 8655 for the full table of the value of the conversion f...

Page 9: ...d in the case of pipettor 2 L also the Seal P If you need to clean other components Unscrew the Shaft Cap M Remove from the Plunger I and pipetting Spring N Remove the Shaft J from the Connecting Bush...

Page 10: ...g Spring N from the shaft Remove the Shaft J from the Connecting Bush L If necessary remove the Seal P from the Plunger I Check the condition of components and clean as recommended Replace any damaged...

Page 11: ...ny damaged parts Assemble the components in reverse order When inserting the shaft or the shaft holder set into the Connecting Bush make sure that they are inserted all the way to the end by gently ro...

Page 12: ...ealing elements e g after repeated autoclaving procedures Sterilization Sterilization in an autoclave The pipettor can be sterilized in an autoclave at 121 C for 20 minutes Sterilization under other c...

Page 13: ...Shaft Kit with seal and O ring 2 SP9811 1 J O Shaft kit with O ring 10 SP9812 1 20 SP9813 1 J Shaft 50 100 SP9814 1 200 SP9815 1 1000 SP9816 1 L Connecting Bush All SP9810 1 H Tip ejector 2 10 SP9817...

Page 14: ...natural forces or other causes not arising from defects in the original material workmanship This warranty does not cover the PTFE seal O ring or shaft Claims for transit damage should be filed with t...

Page 15: ...EUROPE France t 0800 916 882 f 0800 918 636 Germany t 0800 101 1153 f 0800 101 2427 The Netherlands t 20 655 79 28 f 20 659 76 73 United Kingdom t 0800 376 8660 f 0800 279 1117 All Other European Coun...

Page 16: ......

Page 17: ...2019 Corning Incorporated All rights reserved 3 19 CLS AN 525 H EliteTouch IM Corning Incorporated Daniszewska 4 03 230 Warsaw www corning com lifesciences Made in Poland...

Reviews: