Short User's Manual
Gosselin S-Blender 1
Page 2 of 2
(3)
Pull the handle forward to
close the door.
Blending operation starts
automatically.
Wait until set time elapses
and device stops. Message
"End" appears on display.
If you set the time to "HLd",
you have to open the door
to stop the blending
operation.
(4)
Push the handle backward
to open the door.
Take the bag with the
sample out of the chamber.
NOTE: You can open the door during the blending operation, to check the sample. The time on display stops at current
value, and displays are blinking for 10 seconds. During this time, you can check the sample, and if it is not ready, you
can close the door again to continue the blending operation. The time on display starts running again.
If the sample is ready when you open the door during the blending operation, you can wait for 10 seconds and the
displays stop blinking, or you can press any key to cancel the blinking. This means that the run is ended.
NOTE: You can change speed setting during the operation, but you can't change time and program settings.
6. CLEANING
Before starting a cleaning procedure, turn the main switch off and unplug the power cable!
(1)
Push the handle backward
to open the door.
Reach inside the chamber
with your hands and pull
both paddles off the shafts.
Put the paddles in the
waste drawer of the
chamber.
(2)
Lift the door up to unhook it
from the holders on the
bottom side.
(3)
Turn the bottom side of the
door away from the device.
Unhook the top side of the
door by gently pulling it
away from the top holders.
(4)
Pull the chamber away
from the device.
Clean the chamber, the
paddles and the door with
soft cloth and warm water.
For cleaning of the rest of
device use soft damp cloth.
NOTE: To assemble the device, follow the reverse procedure.
7. TROUBLESHOOTING
Following is the table of errors, which could appear on the display of the device, and their descriptions.
ERROR
DESCRIPTION
Err 1
Driver error (driver temperature too high).
Err 2
Motor current too big.
Err 3
Motor is not turning for more than 2 seconds.
Err 4
Motor tries to start 5 times, but it can't because of jamming (too hard or too big sample).
NOTE: Some very hard samples, such as shell-fish, may damage the bags and cause leaks. Make sure to use quality
bags and prepare the appropriate samples.