background image

Short User's Manual 

Gosselin S-Blender 1

Page 1 of 2 

1. INTENDED USE

S-Blender 1 is used for homogenization of the samples in the bags and produces a liquid suspension for further analysis. 

2. OVERALL VIEW

            

3. CONTROL PANEL DESCRIPTION

1  Fast, Normal and Slow indicator lights - for speed indication 

2  SPEED key - for speed setting (Fast, Normal and Slow) 

3  Time display 

4  TIME key - for time setting (in 10 s steps, Min 30 s, Max 210 s, Hold) 

5  PROGRAM SELECT key (0 - 9) 

6  PROGRAM SET key - for entering/exiting the program setting mode 

7  Program setting mode indicator light 

8  Program number display 

9  RUN indicator light 

4. SETTINGS

Plug the power cord to mains socket of the device and to a grounded wall socket. Push the handle backward to open the 

door. Turn on the main switch on the back side of the device. 
Manual settings:

-  Set the speed by pressing the SPEED key. You can set Fast, Normal and Slow speed. 

-  Set the time by pressing the TIME key. You can set the time between 30 and 210 seconds in 10 seconds steps. When 

"HLd" is displayed, Hold function is activated and the time is set to continuous operation. 

Program settings:

- You can select a program (0 - 9) by pressing PROGRAM SELECT key. The preset speed and time are displayed, 

when you select individual programs. 

- You can set the preset speed and time for individual programs. Select the program you wish to change by pressing 

PROGRAM SELECT key, and then press PROGRAM SET key to enter the program setting mode - key indicator light 

turns on. Set the speed by pressing the SPEED key, and time by pressing the TIME key. Press PROGRAM SET key to 

confirm the values and exit the program setting mode - key indicator light turns off. 

NOTE: Short pressing of the keys are used to increase the values. Long pressing of the keys (over 1 second) are used to 

decrease the values. 

5. OPERATION

(1)

Push the handle backward 

to open the door. 

Make sure that waste 

drawer of the chamber is 

empty. 

Set the speed and time 

manually or select a 

program. 

(2)

Insert the bag with the 

sample and diluents into 

the chamber. 

It is recommended to slide 

the bag along the door to 

remove the air out of the 

bag and avoid the leaks 

when closing the door. 

1

2

3

4

5

6

7

8

9

Handle 

Door

Control panel 

Foot 

Waste drawer

Main switch 

Fuse compartment 

Mains socket 

Chamber

Reviews: