background image

6

7

Setting the Volume

After filling the pipet, the end should be dried with an absorbent paper that will not leave 

impurities, in order to remove any excess solution on the outside surface of the pipet. 

To set the required liquid volume precisely, press the dispense button gently (Figure 3C), 

dispense the excessive liquid from the pipet until the meniscus of the liquid aligns exactly 

with the required volume mark on the pipet.

Emptying the Pipet

Holding the vessel in an angled position, place the pipet end in contact with the vessel 

wall and press the dispense button gently (Figure 3C). The dispensing intensity may be 

adjusted depending on pressure applied to the dispense button. The greater the pressure 

applied, the faster the outflow of liquid from the pipet.
The pipet controller has two dispense modes. The dispense mode is selected with the 

BLOW/GRAV switch (Figure 1D).
• Gravity mode marked with the   icon on the display (Figure 2.4): Dispensing is effected 

in gravity mode, which means that the liquid flows out of the pipet by its own weight. 

• Blow out mode marked with the   icon on the display (Figure 2.5): Dispensing is effected 

in gravity mode, however, when the dispense button is pressed to the middle position, 

the pump is started and fast emptying of the pipet with a blow out is effected. 

CAUTION:

 During gravimetric dispensing the pipet may not be completely emptied due 

to the characteristics of pipets used with the pipet controller.

6. Troubleshooting

If during your work the pipet controller is not functioning correctly, check the cause and 

correct the fault.

Problem

Possible cause

Action

The pipet falls out (the 

holding force of the pipet 

is too small), or tilts to the 

side too much.

The pipet holder is dirty or wet 

(Figure 1G).

Take out the pipet holder,  

and clean, wash, and dry it.

The pipet holder is damaged.

Replace the pipet holder 

with a new one.

The pump is working, but 

the pipet controller aid does 

not draw liquid or draws 

liquid very slowly.

The filter is dirty (Figure 1H).

Take out the pipet holder, take out 

the filter, if it is dirty, replace it 

with a new one.

The pipet holder and/or the 

connector gasket is damaged 

(Figure 1J).

Replace the mechanically  

damaged elements with new 

ones.

Liquid leaks from the pipet 

(the aspiration and the 

dispense buttons are not 

pressed)

The pipet is damaged.

Check the pipet for damage 

(cracks, dents) if present, replace 

the pipet with a new one.

The pipet is inserted 

incorrectly.

Check whether the pipet has been 

correctly inserted in the pipet 

holder.

The pipet holder, the filter, or 

the connector gasket is installed 

incorrectly.

Check whether all parts are  

present and correctly installed.

The pipet holder and/or the  

connector gasket is damaged  

(Figures 1G, 1J).

Replace the mechanically  

damaged elements with new 

ones.

If the above actions do not remedy the incorrect functioning of the pipet controller, the 

device should be sent to the nearest authorized service center.

• The pipet controller should not be used for drawing solutions with a temperature above 

50°C. 

• The pipet controller should be operated in the temperature range from +10°C to +35°C.
•  The pipet controller is suitable for general laboratory use only. It must be used only by 

personnel who know the health risks associated with the substances that are normally 

used with this instrument.

4. Turning On the Pipet Controller

The pipet controller is switched on by pressing the aspiration button (Figure 1A) or the 

dispense button (Figure 1B). The display will show the selected aspiration mode, dispense 

mode, and the battery level indicator (Figure 2.3). Examples of the display indications are 

shown in Figure 5. The battery is discharged and requires charging if the indicator displays 

one “bar” (Figure 5.3). When the battery is fully charged, the indicator displays three “bars” 

(Figure 5.1). The pipet controller will switch off automatically if not used for 5 minutes. 
The pipet controller should be charged only with the original charger.
The main voltage should conform with the specification on the charger.
Charging should be done in accordance with section 8 of this instruction for use.

5. Aspirating and Dispensing Liquids

Attaching a pipet 
CAUTION

: Before attaching a pipet, check whether the pipet is not damaged, has no dents 

or sharp edges in the gripping section. Check to ensure the gripping section is dry. 
The pipet controller should be gripped as close to the upper end as possible and carefully 

inserted into the pipet controller holder until resistance is felt (Figure 3A). Do not apply 

excessive force, to prevent damage to thin pipets controller and to avoid injury risk. A pipet 

that has been correctly attached and sealed in the holder should not tilt to the side.
After attaching a pipet, hold the device in such a way as to keep the pipet controller in 

a vertical position with an attached pipet. When not in use, it is not recommended to leave 

the pipet attached for long periods, for example overnight or over a weekend.

Note

: Do not rest the pipet controller on the bench if there is liquid in the pipet.

Filling the pipet

Before aspirating is started, set the suction speed switch using the HIGH/LOW switch 

(Figure 1C):
  • HIGH position: fast aspirating (Figure 2.1)
  • LOW position: slow aspirating (Figure 2.2)
It is recommended to choose the LOW position for pipets with a volume up to 5 mL, and 

the HIGH position for pipets with a volume greater than 5 mL. Holding the pipet controller 

vertically, immerse the pipet end in the liquid to be drawn up (Figure 3B), and press the 

aspiration button gently. The pipet filling speed depends on the pressure applied to the 

aspirating button. The greater the pressure applied, the faster the liquid is aspirated into 

the pipet.
It is recommended to draw a slightly larger liquid volume than required (due to meniscus 

above the required volume mark), reducing the aspiration speed, particularly in the final 

filling stage, so as not to overfill the pipet.

Summary of Contents for Axygen MotoPet

Page 1: ...Axygen MotoPet Pipet Controller Instruction Manual ...

Page 2: ... PP D Dispense mode switch PP E Display F Nose piece PP G Pipet holder SI H Membrane filter PP PTFE J Connector gasket SI K Battery cover PP L Charger 9V EU US UK AU Input 100 240V 50 60Hz 0 3A Output AC DC 9V 230mA Battery type NiHM M Charging base station N Bench stand P Wall mount PP Polypropylene PTFE Polytetrafluoroethylene SI Silicone Casing PP ...

Page 3: ...rom any liability under the warranty Only original spare parts and accessories recommended by the manufacturer shall be used Only the original charger supplied by the manufacturer shall be used for charging the batteries In case of incorrect functioning of the pipet controller work shall be stopped The device shall be cleaned according to section 9 and sent for repair to an authorized service cent...

Page 4: ...roller the device should be sent to the nearest authorized service center The pipet controller should not be used for drawing solutions with a temperature above 50 C The pipet controller should be operated in the temperature range from 10 C to 35 C The pipet controller is suitable for general laboratory use only It must be used only by personnel who know the health risks associated with the substa...

Page 5: ...ten details including the precise specification of solutions used and the type of laboratory in which the device was used should be sent with the product 7 Replacing the filter Note The work safety instructions given in section 2 should be observed when disassembling the pipet controller The filter replacement is necessary if drawing efficiency deterioration is observed The direct reason may be a ...

Page 6: ...uipment WEEE provided with the products and available at the following link www corning com weee The crossed out wheeled bin symbol is printed in the product instruction manual and on the packaging to indicate that in accordance with the requirements of Directive 2006 66 EC of 6th September 2006 on batteries and accumulators and waste batteries and accumulators batteries must be disposed of accord...

Page 7: ... EUROPE France t 0800 916 882 f 0800 918 636 Germany t 0800 101 1153 f 0800 101 2427 The Netherlands t 020 655 79 28 f 020 659 76 73 United Kingdom t 0800 376 8660 f 0800 279 1117 All Other European Countries t 31 0 20 659 60 51 f 31 0 20 659 76 73 LATIN AMERICA grupoLA corning com Brasil t 55 11 3089 7400 Mexico t 52 81 8158 8400 Corning Incorporated Daniszewska 4 03 230 Warsaw www corning com li...

Reviews: