Coretec Trainer 1000 SE User Manual Download Page 6

 
PAGE 6 OF 10  

 USER MANUAL 

 

 

 
 
 
 

 

OPERATING MODES

 

 

Functions 

The Trainer 1000 SE features the same functions as the Trainer 1000: 
 
AED mode (with activatable and de-activatable paddle discharge indicator) 
 
Manual mode (with paddle discharge indicator) 
 
However, the Trainer 1000 SE also features an 
 
ECG monitoring mode (requires Upgrade Kit, 3-lead ECG cable, only in monitoring mode) 
 
which can be used to display a range of ready-installed ECG rhythm recordings on the display and to match the 
rhythm shown on the display to the simulated patient’s condition using the remote control. 
 
 
 
 

AED MODE 
 

As with real automated external defibrillators, the 

 evaluates the selected ECG rhythms and issues 

specific instructions when it identifies a shockable rhythm. The delivery of shocks (defibrillation) to the simulated 
patient requires the active involvement of the person operating the defibrillator. 
 
 

MANUAL MODE  
 

This mode allows expert users to decide on administering defibrillation shocks on the basis of the ECG analysis 
derived through the electrodes. Instructors can chose the ECG recording to be shown on the display to provide 
advanced training in the use of additional and drug-based treatment. The heart rhythm can be changed at any 
time in manual mode. If the Trainer 1000 is configured in such a way that the analysis function can be used in 
manual mode, users will also be able to use the AHA and ILCOR treatment -recommendations-based computer-
assisted decision-making aid regarding whether or not to defibrillate.  
 
 

ECG MONITORING MODE 
 

In ECG mode, the electrical activity of a patient’s heart can be monitored using conventional ECG electrodes that 
are attached using snap fasteners. The readout provided on the screen represents the consolidated data from lead 
II. The instructor can select one of the rhythm recordings stored on the system by remote control and adjust the 
recording to the simulated patient condition. The device does not give voice prompts. If the device identifies a 
heart rhythm that potentially calls for defibrillation, it will prompt the user to check the patient and to change to 
the defibrillation electrodes.  
 

Summary of Contents for Trainer 1000 SE

Page 1: ...SE Supplementary user manual ONLY FOR USE IN CONJUNCTION WITH THE TRAINER 1000 USER MANUAL V 1 0 GBR ...

Page 2: ... and training device for realistic Advanced Life Support training on a defibrillator AED and providing treatment to patients on the basis of an ECG diagnosis Due to the similarity of its design and operating elements the Trainer 1000 is particularly suitable for training and instructing people in the use of the Lifepak 1000 series defibrillators The Trainer 1000 does not deliver energy through adh...

Page 3: ...ersonnel that is not expressly and legally authorised to perform electrotherapy in the country of application are advised to consult the head of their department before use IMPORTANT INFORMATION ON PERFORMING ECGS WITH THE TRAINER 1000 The LIFEPACK 1000 SE s monitoring function is designed to monitor the heart s electrical activity through a stable limb lead II cable The lead and the amplitude set...

Page 4: ...to the device that the monitoring electrodes have been attached and makes it start showing a heart rhythm NSR SBrady STachy Pressing this button will cause all shock decisions to be changed to NO SHOCK ADVISED This function can only be used if either the messages PRESS FOR ANALYSIS or ANALYSIS IN PROGRESS STEP BACK are being displayed Pressing this button will bring up a normal sinus rhythm a sinu...

Page 5: ...d to turn off the AED Trainer Pressing this button will bring up a bigeminus a 3 1 AV block or atrail fibrillation on the display To change to ECG mode Pressing this button will make the Trainer 1000 only use the signals from the connected lead for monitoring Pressing this button will make the device switch to ECG mode No function Pressing this button will enable you to activate the heart rhythms ...

Page 6: ...hocks on the basis of the ECG analysis derived through the electrodes Instructors can chose the ECG recording to be shown on the display to provide advanced training in the use of additional and drug based treatment The heart rhythm can be changed at any time in manual mode If the Trainer 1000 is configured in such a way that the analysis function can be used in manual mode users will also be able...

Page 7: ...ECG will now be shown On the remote control ECGs can be selected as follows Black ECGs Press button Red ECGs Press the red button first and then the ECG button Blue ECGs Press the blue button first and then the ECG button To exit ECG mode simply reverse these steps If the rhythm shown by the ECG is shockable the Trainer 1000 will prompt you to connect defibrillation electrodes and to exit ECG mode...

Page 8: ...n shockable rhythm Based on non shockable rhythm Fix 1 SSSSK FFFFSr Fix1 Fix2 Fix2 KKKK AAAA Fix1 Fix2 Fix3 SSKKKK FFAASrSr Fix3 Fix5 Fix4 SSKSSK FFAFFSr Fix4 Fix5 Fix5 KSKSS AFSrFF Fix4 Fix5 Customer XXXXX S orr K F oder A Customer Fix4 Customer Fix5 S Shock advised K No Shock Advised Sr Sinus rhythm F Fibrillation A Asystole ...

Page 9: ...able The designations of the ECG leads leading to the patient can vary but are generally as follows AHA designations IEC designations Electrode position RA R Right arm LA L Left arm LL F Left leg If your mannequin does not have any connections for the ECG cable you can use conventional adhesive ECG electrodes In this case we recommend using Ag AgCl solid gel adhesive electrodes and to leave them o...

Page 10: ...ctrodes has been properly inserted into the socket on the AED Trainer and the instructor has verified that the electrodes have been correctly attached via the remote control If while using the AED Trainer the training electrodes are removed from the device it will display the message CONNECT ELECTRODES the training electrodes on the other hand come off the mannequin it will not display this messag...

Reviews: