Declaration!

This user manual is a integrated user manual of the E51.B3S series piezoelectric controller. 
Please read this user manual carefully before using this controller. Follow the instructions in the 
manual during use. If there is any problem, please contact us for technical support. If you do not 
follow this manual or disassemble and modify the product yourself, the company will not be liable 
for any consequences arising therefrom.
Please read the following to avoid personal injury and to prevent damage to this product or any 
other product connected to it. In order to avoid possible hazards, this product can only be used 

within the specified range.

Notice!

Do not touch any exposed ends of the product and its accessories.
There is high voltage inside. Do not open the case without permission.
Do not connect or disconnect input, output, or sensor cables with power on.
Please keep surface of E51.B3S clean and dry, don't operate in humid or static environment.

After use, output voltage should be cleared to zero before turning off the controller switch, such as 

switching the servo state to the open-loop state.

Danger!

The piezoelectric power amplifier described in this manual is a high-voltage device capable of 

outputting high currents, which can cause serious or even fatal damage if not used properly.
It is strongly recommended that you do not touch any parts that connect to the high voltage output.
Special Note: If you connect it with other products in addition to our company, please follow the 
general accident prevention procedures.

Operating the high-voltage amplification requires training professional operators.

Cautious!

Warning!

If the voltage exceeds the PZT's tolerable range, it will cause permanent damage to the PZT. 
Before adding voltage to the PZT poles, it must be ensured that the positive and negative poles 
of the PZT are connected correctly and the operating voltage is within the allowable range of this 
PZT.

E51.B3S housing should be installed on a horizontal surface in an area with a 3CM air flow area to 

prevent internal convection in the vertical direction.

Insufficient airflow can cause equipment to overheat or premature instrument damage.

DECLARATION

Summary of Contents for E51.B3S Series

Page 1: ...This document describes the following products E51 B3S H1 Piezo Controller SGS sensor E51 B3S I1 Piezo Controller SGS sensor E51 B3S Series Piezo Controller User Manual Version V1 0...

Page 2: ...uld be cleared to zero before turning off the controller switch such as switching the servo state to the open loop state Danger The piezoelectric power amplifier described in this manual is a high vol...

Page 3: ...2 Applicaitons 3 2 3 Order information 3 3 Driving Principle 4 4 Power Calculation 4 5 Panel Introduction 5 6 Notes and suggestions 6 7 Maintenance Storage Transportation 6 7 1 Cleaning measures 6 7 2...

Page 4: ...incorrectly there will be a possibility of leakage If you touch the E51 B3S piezo controller it may cause serious or even fatal injuries If the piezo controller housing is opened without permission to...

Page 5: ...uld install operate maintain and clean the E51 B3S 2 Introduction 2 1 Features 3 channels small size 20 30V power supply Peak current 180mA Ave current 25mA Unload bandwidth 1KHz Output short circuit...

Page 6: ...ive voltage V Vs Vs Cpiezo Piezo actuator capacitance F f Operating frequency of the sine wave Hz 3 Driving Principle Analog input Control input Open loop Servo Sensor servo control Sensor output Sens...

Page 7: ...r supply ground 10 Sensing power supply ground 11 Drive output ground 12 CH 2 sensor input 13 CH 2 sensor input 14 CH 1 sensor input 15 CH 1 sensor input No Pin definition 1 24V power supply ground 2...

Page 8: ...ing that any type of conductive particles metal dust or debris pencil lead screws enter the device Be careful not to drop the equipment when cleaning to avoid any form of mechanical shock Disconnect t...

Page 9: ...ice E51 B3S is a precision instrument which should be handled with care In case of any problem please record the problem and contact CoreMorrow to be repaired by professional technicians If the instru...

Page 10: ...Contact us Harbin Core Tomorrow Science Technology Co Ltd Tel 86 451 86268790 Email info coremorrow com Website www coremorrow com Address Building I2 No 191 Xuefu Road Nangang District Harbin HLJ Chi...

Reviews: