Planar
TR1300/1,
TR5048,
TR7530
Programming
Manual
SCPI
202
SENS:SWE:TYPE
SENSe<Ch>:SWEep:TYPE
{LINear|LOGarithmic|SEGMent|POWer}
SENSe<Ch>:SWEep:TYPE?
Description
Sets
or
reads
out
the
sweep
type
(command/query).
Target
Channel
<Ch>,
<Ch>={[1]|2|3|4|5|6|7|8|9}
Parameter
"LINear"
:
Linear
frequency
sweep
"LOGarithmic"
:
Logarithmic
frequency
sweep
"SEGMent"
:
Segment
frequency
sweep
"POWer
"
:
Power
sweep
Out
of
Range
Error
occurs.
The
command
is
ignored.
Error
code:
206.
Query
Response
{LIN|LOG|SEGM|POW}
Preset
Value
LIN
Equivalent
Softkeys
Stimulus
>
Sweep
Type
SERV:CHAN:ACT?
SERVice:CHANnel:ACTive?
Description
Reads
out
the
active
channel
number
(query
only).
Query
Response
<numeric>
from
1
to
16
Equivalent
Softkeys
None