Cooper DCM 495 User Manual Download Page 5

CF 52A 

Mar 2014 

 

 

Chapter 3 - Calibration 

3.1 Calibration 

The following calibration process is used to fully calibrate the DCM 495 with the transducer being used. 
 
Connect the DCM 495 to the transducer observing the wiring codes as explained in Chapter 2. Apply 
power and allow the unit to stabilize for 10-15 minutes before starting the calibration process. 
 
ZERO ADJUSTMENT: With no load or pressure on the transducer and using the ZERO potentiometers, 
adjust the output on the readout instrument to Zero. 
 
SPAN ADJUSTMENT: Load the transducer to its full scale and using the SPAN potentiometer, adjust the 
span to indicate the full scale reading on the readout instrument. 
 
Repeat this process to ensure that calibration is completed. 

3.2 Using Shunt Calibration 

The  Cooper  DCM  495  features  shunt  calibration. This  calibration  method  simulates  an  output from  the 
transducer as if a load or pressure is being applied to it. This is accomplished by placing a known resistance 
across one leg of the Wheatstone bridge. When performing shunt calibration, the transducer should be at 
ZERO pressure or load (in other words, no physical load or pressure is to be applied to the transducer). 
The full scale output and shunt cal output is found on the TRANSDUCER CALIBRATION SHEET. This 
information  can  be 

used  to  calibrate  the  amplifier’s  output  voltage  and  the  readout’s  display  with  the 

following equations: 
 

 

The following formulas are used to calculate output voltage and display units: 

 

 

Reviews: