CF 52A 

Mar 2014 

 

Step 3 

Locate the full scale output in mV/V for the transducer on the calibration certificate. 

 
Step 4 

From the Coarse Gain Table (Figure 2-3) below, find the nearest sensitivity to that of the 

transducer and put the corresponding switch or switches to the ON position. 
 
This is the initial setup that will allow you to fully calibrate the DCM 495 with the transducer. 
The calibration procedure is explained in Chapter 3. 

 

2.3 Panel Mounting Instructions 

The DCM 495 In-Line Amplifier can be easily mounted to a panel by using the measurements below. The 
cover must be removed to have access to the mounting holes. #6 or #8 screws should be used for mounting 
the box to a panel. 

 

 

 

Coarse Gain Setup 

 

Figure 2-3 