Prep 1
Prep 2
Remove device from package and attach
catheter to handle, ensuring connection
is secure.
Activate CO
2
cartridge by turning
red
activation
knob until it stops.
Step 1
Before inserting catheter into accessory
channel, identify bleeding site, remove as
much blood as possible, then flush accessory
channel with air.
Step 2
Slowly advance catheter through accessory
channel in short increments until catheter tip
is visualized endoscopically.
Precaution:
To
avoid catheter occlusion, do not place catheter
directly in contact with blood and/or mucosa,
including any pooled blood and do not aspirate
blood while catheter is in accessory channel. To
ensure proper visibility, catheter tip should be
1-2 cm away from bleeding site at all times.
Hemospray
®
E N D O S C O P I C H E M O S TAT
QUICK REFERENCE GUIDE
Step 3
To allow powder deployment, turn
red
valve to
open position.
Note:
Device is now active and
ready for use. Do not press trigger button until
powder deployment is desired.
M E D I C A L