COOK Medical Hemospray Quick Reference Manual Download Page 2

Step 6

Upon completion of procedure, depressurize 
device by rotating 

red 

activation knob 

counterclockwise until CO

2

 cartridge 

depressurizes completely.

Step 5

Prior to removing Hemospray device from 
patient, turn 

red

 valve to closed position. To 

remove device from endoscope, grasp catheter 
and pull slowly.

Dispose of device per institutional guidelines 
for biohazardous medical waste. 

Refer to current instructions for detailed system use.

Summary of Clinical Data:  

Clinical data summary information that was, 
in part, the basis for granting the de novo 
can be found on the Cook Medical website at 

CookMedical.com/HemosprayData

Step 4

To deploy powder, depress 

red

 trigger 

button for 1-2 seconds and release. Continue 
applying in short bursts until site is completely 
covered with powder and no active bleeding 
is visualized. Multiple applications may be 
required; however, no more than (3) Hemospray 
devices should be applied per patient.  

Note:

 Allow time for hemostasis before 

aspirating. Irrigation of treated site is not 
required after powder deployment. If irrigation 
is desired, allow time for hemostasis before 
irrigating with water. To avoid potential 
endoscope occlusion, do not aspirate powder 
into endoscope channel. If catheter becomes 

occluded during procedure, turn 

red

 valve 

to closed position, remove catheter from 
endoscope and replace with extra catheter 
provided in package. Then repeat steps 1-4.  
If more powder is needed and spray canister 
and/or CO

2

 cartridge is empty, a new device  

is required.

M E D I C A L

© COOK 05/2018  ESC-D43164-EN-F

AI, ESC, IR, OHNS, PI, RH, SUR-8.5X11