1
May 2013 |
257775R00
CONVIRON
CMP 6050 RETROFIT INSTALLATION INSTRUCTIONS
CMP6050 Retrofit Installation Instructions
Required Parts and Materials:
F
Control panel assembly
F
Retrofit cover panel
F
Control panel bezel assembly
F
Control driver boards
F
Control driver board cables
F
New/replacement sensors
F
Pilaster and light sensor support bracket
F
Retrofit Hardware Package
F
Spiral wire protection wrap
F
Zip ties
F
Tool kit appropriate for servicing Conviron
chambers
F
Anti-static wrist strap
F
Schematic #231106: SCHEM-Communicaion
Wiring CMP6050
F
Schematic #257349: SCHEM-Main Control,
Retro CMP6050 for 4000/5000, 50/60hz
F
CMP6050 Operator’s Manual
1.
Record information programmed into the controller being replaced prior to beginning installation,
including:
i.
Saved programs
ii.
Schedule
iii.
Network settings
iv.
Alarm Limits
2.
Disable power to the chamber by turning OFF the main breaker supplying power from building
electrical service. If the chamber is equipped with a Uninterruptable Power Supply (UPS), ensure it
is also OFF. Switch all breakers in the control panel to the OFF position.
3.
Disconnect all cables and wires running from the control panel to the old controller being removed.
Disconnect at the terminal blocks within the control panel.
F
F
F
Procedure: