Default alarm settings 

When the unit is turned on, the alarms automatically default to out-of-range set points so 
that the alarms will not trigger. The default settings are: 

Relative Humidity 

HI 

101.00 

LO 

–01.00 

Temperature °F 

HI 

221.00 

LO 

0.00 

Temperature °C 

HI 

105.00 

LO 

–17.77 

Dew Point °F 

HI 

141.00 

LO 

0.00 

Dew Point °C 

HI 

60.56 

LO 

–17.77 

To prevent any alarm from tripping, simply set the alarm set point out of the unit’s range. 
As an example, in relative humidity any alarm set point above 100 or less than zero will 
never trip the alarm. To quickly reset all alarm values to the default out-of-range 
condition, simply turn the unit off, wait 2 seconds, and then turn it on again. 

Default alarm sounds 

Three distinctive alarm sounds distinguish each measurement. The alarm sounds are: 

Dew Point 

1 beep per second 

Temperature 

2 beeps per second 

Relative Humidity 

3 beeps per second 

 

Note:

  If two or more alarms are sounding simultaneously the sound will be heard in the 

following sequence:  dew point, temperature, and relative humidity.  

 

GENERAL INFORMATION 

Display 

For ease of use the display is optimized at a 20-degree angle. It is best viewed by the 
user when sitting or standing with the unit flat on a lab bench or desk. When viewed 
straight-on the display will appear to have less contrast. 

Key sound 

An audible confirmation beep will be heard each time a button is pressed. 

Freezing points 

Dew points below freezing should be converted to frost points by using frost point 
tables. 

USER CALIBRATION 

WARNING:

  User calibration adjustment will void the factory-supplied National Institute 

of Standards and Technology (NIST) Traceable

®

 Certificate. The unit has been 

calibrated against NIST traceable instrumentation. 

For users wishing to check factory calibration, we suggest using a calibration laboratory 
with capabilities to generate a known humidity and utilize sophisticated chilled mirror 
test equipment. 

Salt as a calibration standard 

We do NOT recommend or use the saturated salt calibration method. NIST does NOT 
use the salt method. The proper technique involved in the salt method is far more 
difficult and far more time consuming than most literature suggests. Liquid volume, air 
volume, sensor placement, temperature, air movement, vapor pressure, equilibration 
time, and a host of other factors can all contribute to gross errors. On occasion these 
errors are reproducible. 
For information on the saturated salt method we suggest studying American Society of 
Testing Materials Method ASTM E104 and obtaining a paper from the National Institute 
of Standards and Technology JOURNAL OF RESEARCH, Vol. 81A, No. 1, pages 89–
96. 

Dial hygrometers as a calibration standard 

Absolutely do NOT calibrate against hair dial hygrometers. Inaccuracies in dial units 
caused by age of hair and other factors can greatly affect readings. Dial units are not 
sensitive to rapid changes and readings generally lag actual humidity changes by hours. 
Users of some dial units have encountered errors of over 50% relative humidity. 

Sling hygrometer calibration 

Absolutely do NOT calibrate against sling units. Sling units are limited to the accuracy 
and resolution of their thermometers. Temperature errors of as little as 0.05° can cause 
significant humidity errors. Unfortunately temperature, time, air volume, and technique 
can cause reproducible, but large errors. In low humidities, in small spaces, the addition 
of a wet wick may change the actual humidity. If too much swing time is taken, a 
partially dry wick produces errors. If too little swing time is taken, sufficient evaporation 
may not have taken place to cool the thermometer. Sling units generally measure crude 
averages of a large volume of air. Users of some sling units have encountered errors of 
over 50% relative humidity. 

 

 

 

 

 

 

ACCESSORIES AVAILABLE 

Cat. No. 1023 AC Adaptor for continuous use 

An accessory adaptor is available for continuous AC operation. When using 
the adaptor, the non-rechargeable lithium or alkaline battery must be removed. 
Never use the adaptor with a non-rechargeable battery. A battery is NOT 
required to use the adaptor. A rechargeable NiCad battery may be used with 
the adaptor. The adaptor will simultaneously power the unit and recharge a 
NiCad battery. 

Cat. No. 4020 Rechargeable NiCad Battery 

Requires use of Cat. No. 1023 AC Adaptor for continuous use. 

Cat. No. 4099 Data Acquisition System Accessory 

Powerful and easy to use computer data capture/data logging program works 
with Traceable

®

 Instruments with computer output. Records interval readings 

from 1 to 10,000 seconds; displays minimum/maximum readings; and utilizes 
an alarm mode that permits the user to be notified visually, audibly, and by 
email when an alarm is triggered. Data is stored to a file that can be printed in 
any report or spreadsheet format. Networking server/client capability allows 
the captured data to be monitored on a remote workstation and/or by email. It 
is designed to work with Windows

®

 98/Me/NT/2000/XP/Vista.  Includes a CD, a 

6-foot cable (supplied USB and serial connections) that plugs into the 
instrument and computer. Accessory extension cables expand cable length to 
300 feet. 

 

 

ALL OPERATIONAL DIFFICULTIES 

If this meter does not function properly for any reason, please replace the battery with a 
new high quality battery (see “Battery Replacement” section). Low battery power can 
occasionally cause any number of “apparent” operational difficulties. Replacing the 
battery with a new fresh battery will solve most difficulties. 
 

Frozen display 

Turning the unit on and off rapidly may lock (freeze) the display. Wait 2 seconds before 
switching the unit from off to on. If a “frozen display” occurs, remove the battery, wait 1 
minute, insert the battery, and resume operation. 

 

BATTERY REPLACEMENT 

A faint display or no display are indicators that the battery must be replaced. A BAT 
indicator appears on the display when the battery is low. Remove the battery cover on 
back of the unit. Remove the exhausted battery and replace it with a new 9-volt alkaline 
or lithium battery. Replace the battery cover. Do NOT use a heavy-duty or regular 
battery since they do not have sufficient power. If BAT indicator appears on the display 
when using a NiCad battery, recharge the battery. 

 

Cat. No. 4080 

Traceable

®

 Humidity/Temperature/Dew Point Meter (Universal) 

Cat. No. 4083 

Traceable

®

 Humidity/Temperature/Dew Point Meter (Basic) 

Cat. No. 4085 

Traceable

®

 Humidity/Temperature/Dew Point Meter (Fast-Response) 

 
 
 
 
 
 
 
 
 
 
 
WARRANTY, SERVICE, OR RECALIBRATION 

For warranty, service, or recalibration, contact: 

CONTROL COMPANY 

4455 Rex Road 

Friendswood, Texas 77546 USA 

Ph. 281-482-1714   Fax 281-482-9448 

E-mail [email protected] 

www.control3.com 

 

 

Control Company is ISO 9001 Quality-

Certified by DNV and ISO 17025 accredited 

as a Calibration Laboratory by A2LA. 