- 29 -
Ambient Temperature
The ambient temperature is decided from the multiple measurement points which are a 50mm-
distance and a 100mm-distance from the product.
During the operation, adjust the air current to make certain that the temperatures measured in the
points stay within the specified temperature.
CAUTION
Note that although the ambient temperature is within the specified range, an operational
malfunction may occur if there is other device generating high heat; the radiation will influence
the product to increase its temperature.
Measurement points of ambient temperature
50mm or more
50mm or more
50mm or more
50mm or more
50mm or more
50mm or more
50
50
50
[mm]