ECG1200G Electrocardiograph - User Manual 

 

 

2

1.2.23 Net Weight: 3.2Kg 

Chapter2 Safety Notes

 

2.1 The power supply should be grounded properly before operation. 

2.2 If there are any questions for the integrality of protective grounding cable, the device must be 

run with built-in power supply. 

2.3 Please pull out power supply plug before changing the fuse. 
2.4  This  device  must  be  operated  by  medical  staff  trained  technically  and  professionally, 

preserved by special person. 

2.5  The  operator  must  read  this  instruction  manual  carefully  before  operation,  and  operate  the 

device according to operation regulations strictly. 

2.6  The  design  of  this  device  has  mature  consideration  of  security,  but  operator  should  never 

neglect attention to device state and patient’s observation. 

2.7 Please turn off the device and pull out power supply plug before cleaning and disinfection . 
2.8 Please don’t use this device in the presence of flammable anaesthesia gas. 
2.9 Don't rub the screen with edge tools or sharp materials. 
2.10  If  this  device  is  used  with  cardiac  defibrillator  or  other  electric  stimulating  devices  at  the 

same  time,  please  choose  Ag/AgCl  chloride  chest  electrode  and  ECG  lead  cables  with 
defibrillation function. To prevent the metal electrode from burning patients' skin, the disposable 
chest electrode should be used if the defibrillation time is over 5seconds.  It is better not to use 
this device with other electric stimulating devices at the same time. If it is necessary, there must 
be professional technician guiding on the scene. 

2.11  When  connected  with  this  ECG  device,  others  must  be  Type  I  equipment  complied  with 
IEC60601-1.  Because  the  total  leakage  current  may  injure  patient,  the  monitoring  of  leakage 
current shall be carried out and taken charge by the connected devices. 
2.12 Notes concerning ECG waveform measurement and analysis 
(1)

 

P  wave  and  Q  wave  identify  are  not  always  reliable  with  intensive  EMG  or  AC 
interference. Neither are the ST segment and T wave with baseline drift.   

(2)

 

Winding and unclear end position of S wave and T wave may cause error in measurement.   

(3)

 

When  R  wave  is  uninspected  caused  by  some  leads  off  or  QRS  wave  low  voltage,  the 
heart rate measurement may deviate greatly from the correct. 

(4)

 

In case of QRS low voltage, ECG axis calculation and border-point identify of QRS wave 
are not always reliable. 

(5)

 

Occasionally,  frequent  ventricular  premature  complexes  may  be  identified  as  dominant 
beat.   

(6)

 

Merging  of  versatile  arrhythmias  may  result  in  unreliable  measurement  because  of  the 
difficulty in distinguishing P wave in such situation.   

(7)

 

This  device  is  designed  with  auto  analysis  function,  which  only  analyses  the  ECG 
waveforms it gathers and does not reflect all patient's states. Its analysis results may be not 
in  accordance  with  doctor's  diagnoses.  Therefore,  the  final  conclusion  concerning  each 
patient  is  up  to  the  doctor  basing  on  analysis  results  ,  patient  symptoms,  and  other 
examinations together. 

Summary of Contents for ECG1200G

Page 1: ...the manual without our company s written permission is prohibited All information contained in this publication is believed to be correct Our company shall not be liable for incidental and consequent...

Page 2: ...ration 11 Chapter8 Notes During Operation 12 Chapter9 Usage of Recording Paper 13 Chapter10 Electrode Placement 14 10 1 Chest Electrode Placement 14 10 2 Limb Electrode Attachment 14 10 3 Check List f...

Page 3: ...hecking Electrodes Placement 30 13 12 About Us 31 Chapter14 Troubleshooting 33 14 1 Powering off Automatically 33 14 2 AC Interference 33 14 3 EMG Interference 33 14 4 Baseline Drift 33 14 5 Troublesh...

Page 4: ...ead Standard 12 leads model BIP0057 1 2 4 Patient leak current 10 A 1 2 5 Input impedance 50M 1 2 6 Frequency response 0 05Hz 150Hz 3dB 0 4dB 1 2 7 Time constant Time constant 3 2s 1 2 8 CMRR 60dB 100...

Page 5: ...tter not to use this device with other electric stimulating devices at the same time If it is necessary there must be professional technician guiding on the scene 2 11 When connected with this ECG dev...

Page 6: ...operation notes 3 3 2 Faults or injuries caused by dropping accidently when moving after purchasing 3 3 3 Faults or injuries caused by repair reconstruction decomposition etc not in our company 3 3 4...

Page 7: ...e button operation to complete printing and storing 4 13 The device adopts digital signal processor through AC DFT and EMG filters to obtain electrocardiogram of high quality 4 14 The device has such...

Page 8: ...1200G Electrocardiograph User Manual 5 PATIENT Lead connector WEEE 2002 96 EC This item is compliant with Medical Device Directive 93 42 EEC of June 14 1993 a directive of the European Economic Commun...

Page 9: ...ph User Manual 6 Chapter5 Electrocardiograph Panel Sketch Map 5 1 Each Panel View and its Components Figure 5 1 Front view Figure 5 2 Side view 1 Lead Connector 2 USB Port 1 2 3 4 5 Display Screen Pap...

Page 10: ...d to turn the device on or off Function key This key is used to begin or stop sampling Function key This key is used to select filter function Function key This key is used to adjust paper speed Funct...

Page 11: ...ital and Small letter Character key 0 Space Number Capital and Small letter Character key 1 Number Capital and Small letter Character key 2 abc ABC Number Capital and Small letter Character key 3 def...

Page 12: ...umber Capital and Small letter Character key 9 wxyz WXYZ 5 3 Indicator Light Definition When green the light indicates that the device is powered by AC supply while red and green it indicates the batt...

Page 13: ...the places with high air pressure temperature and humidity beyond common standard bad ventilation excessive dust or salt alkali gas and chemical medicine 6 3 The device should be put on flat place an...

Page 14: ...Ensure the electrodes directly connected with patient are placed correct and reliable 7 3 If you have selected to purchase UPS check whether the output voltage is normal or not 7 4 Electric gel coats...

Page 15: ...er using 8 5 Disconnect the power and remove the ECG lead cables gently without strong force 8 6 Properly keep the device and its accessories for operation next time 8 7 Recording Paper Loading 8 7 1...

Page 16: ...ce s worn up and shortening the service cycle of such important components as printer head Please consult your manufacture or our company for purchasing this recording paper 9 3 High temperature humid...

Page 17: ...lary line at the horizontal lever of V4 Clean the skin where chest electrodes are to be attached with alcohol then apply ECG cream to here around 25mm in diameter and to the edge of chest electrodes t...

Page 18: ...er it is inserted to the ECG connector 10 3 Check List for Electrodes and ECG cables Electrode Location Electrode Symbol Plug No Right Arm RA R 9 Left Arm LA L l0 Left Leg LL F 11 Right Leg RL N 14 Ch...

Page 19: ...power supply is off plug the 3 line connector in the device the other end into power receptacle which should be center grounded It is prohibited to use water pipe or other pipeline as grounding cable...

Page 20: ...will turn off automatically to avoid forever damage to the battery 12 3 The battery should be recharged in time after discharged completely If not used for long the battery should be recharged every...

Page 21: ...ction Buttons Click to Enter the sampling interface when powered on the device will automatically enter this interface Click to enter case management interface where you can query modify or delete cas...

Page 22: ...keyboard to enter sampling interface Note Because of the setup about information input time in system settings it needs to input case information before sampling refer to 13 3 Patient Information Inpu...

Page 23: ...sh Printing With the printing button on the keyboard you can begin or finish the printing operation Auto Mode After starting printing automatically the system prints and stores synchronous 12 leads wa...

Page 24: ...ws Here click Caps button to switch between number capital letter and number small letter press Space to input a space press Bkspace to delete the character input last press OK to confirm input and ex...

Page 25: ...interface as shown below See Figure 13 6 Figure 13 6 This interface displays all cases stored in the device The user can select the case needed using the querying function Refer to 13 5Case Querying...

Page 26: ...better to input all decided querying conditions select Cond And to find the corresponding case quickly 13 6 Case Review In case management interface select the case you want to review click Review to...

Page 27: ...Figure 13 9 In this interface user can switch printing modes by the button click to begin printing 13 7 Time and Data Settings In the main interface select button and data and time setting dialogue bo...

Page 28: ...anual 25 Figure 13 10 In this interface user can switch options through and use to edit option content 13 8 System Settings In the main interface select button and system settings dialogue box shown b...

Page 29: ...degree After setting light degree the screen will display different backlight strength Auto off 1 minute 3 minutes 5 minutes 10 minutes 15 minutes 30 minutes 60 minutes None If there is no operation f...

Page 30: ...mpling settings will go back to the default Each item its options and explanation are shown in the table below Item Options Explanation AC Filter ON OFF Set up whether to use AC Filter or not EMG Filt...

Page 31: ...re beat Pause Time ms 1200 3000 The system will use the input value as a standard of judging beat pause Tachycardia bpm 0 250 The system will use the input value as a standard of judging tachycardia B...

Page 32: ...each row waveform Avera ge QRS 3 4 Mark 3 4 None In printing mode when selecting Rhythm Or Auto the system will use the format selected to print average QRS waveform Auto Diag All Data Conclusion Non...

Page 33: ...ata to the data before pressing Print key or after pressing the key Rhythm 1 I II III avR avL avF v1 v2 v3 v4 v5 v6 Set the rhythm lead used for rhythm patterns lead print Rhythm 2 I II III avR avL av...

Page 34: ...placement shown below See Figure 13 14 Figure 13 15 Click std 12 to switch the placement sketch map about 12 leads and add 6leads Click return to exit interface 13 12 About Us Select the button in the...

Page 35: ...ECG1200G Electrocardiograph User Manual 32 Figure 13 16 This interface shows the device name version number company name copyright and our contact details...

Page 36: ...rference Whether or not the device is grounded reliably Whether or not the electrodes or lead cables are connected correctly Whether or not the electrodes and skin are smeared enough electric gel Whet...

Page 37: ...bly 3 There is AC interference 4 Patient is nervous and can not keep quiet 1 Check lead cables grounding cable and power 2 Do patient treatment well Baseline burr 1 AC interference is large 2 Patient...

Page 38: ...35 Unclear waveform 1 The printer head surface is dirty 2 The thermal paper problem 1 In case of power cut clean the printer head with alcohol do not begin printing until the head is dry 2 Replace the...

Page 39: ...tenance and future upgrades to this device must be carried out by personnel trained and authorized by our company The repair should be for our company s original components only 15 2 Please pull out t...

Page 40: ...V air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 1 Guidance and manufacturer s decla...

Page 41: ...in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for...

Page 42: ...MHz 3 V m 80 MHz to 2 5 GHz 3 V 3 V m Portable and mobile RF communications equipment should be used no closer to any part of the model ECG1200G Electrocardiograph ECG including cables than the recomm...

Page 43: ...ING Recommended separation distances between portable and mobile RF communications equipment and the ECG 1101 Series ECG The model ECG1200G Electrocardiograph ECG is intended for use in an electromagn...

Page 44: ...nts contact the local authorities the local recycling center or the shop where the product was purchased If the equipment is not disposed of correctly fines or penalties may be applied in accordance w...

Page 45: ...tection waste electrical products should not be disposed of with household waste Please recycle where facilities exist Check with your local Authority or retailer for recycling advice Symbol for manuf...