ECG1200G Electrocardiograph - User Manual
28
Show Gain
[2.5mm/mV]/[5mm/mV]/[10m
m/mV]/[20mm/mV]
/[40mm/mV]
Set up the ECG gain in screen.
Show Speed
[5mm/s]/[6.25mm/s]/[10mm/s]/[
12.5mm/s]/[25mm/s]/[50mm/s]
Set up ECG sweep speed in screen, but
when printing in auto and rhythm
mode , it does not support 5mm/s,
10mm/s, 12.5mm/s.
Premature(%)
0 ~ 100
The system will use the input value as a
standard of judging premature beat.
Pause Time(ms)
1200 ~ 3000
The system will use the input value as a
standard of judging beat pause.
Tachycardia(bpm
)
0 ~ 250
The system will use the input value as a
standard of judging tachycardia.
Bradycardia(bpm
)
0~99
The system will use the input value as a
standard of judging bradycardia.
13.10 Print Settings
Select the
button in the main interface, the printing setting dialogue box will pop up
shown below.(See Figure 13-13)
Figure 13-13
Click the button[Default], the print settings will go back to the default.
Each item, its options and explanation are shown in the table below.