Chapter 18
NIBP SPECIFICATION
Name
Electronic Sphygmomanometer
Display mode
2.8'' color LCD Display
The degree of protection against ingress of water
IPX0
NIBP specification
Measurement method
Oscillometric method
Working mode
Automatic
Operation mode
Continuous operation
Measurement Range
Pressure:0~290 mmHg (0~38.6 kPa)
Pulse: 40~240/min
Resolution
Pressure: 1mmHg(0.133 kPa)
Accuracy
Static pressure: ±3 mmHg (±0.4 kPa)
Error
The BP Value of the device is equivalence with the measurement
value of Stethoscopy. The error meets the requests in the ANSI/
AAMI SP-10:2002+A1:2003 +A2:2006
Operating temperature/ humidity
+5 ºC~40 ºC 15%RH~85%RH (Non-condensing)
Transport
Transport by general vehicle or according to the order contract,
avoid pounded, shake and splash by rain and snow in transpor-
tation.
Storage
Temperature: -20 ºC~+55 ºC;
Relative humidity: ≤95%; No corrosive gas and drafty.