•  The claim for SpO

2

 accuracy should be supported by clinical research measurements covering the entire spectrum. 

By artificially inducing to different stable oxygen levels, make it in the range of 70% ~ 100% of SpO

2

. Use secondary 

standard SpO

2

 measuring equipment for comparison to collect SpO

2

 values together with the tested product, compose 

paired data groups for accuracy analysis.

•  When using the device, please keep it away from the instruments that can generate strong electric or magnetic field. 

Use the device in an inappropriate environment may cause interference to the surrounding radio equipment or affect its 

working.

•  If necessary, please log on our company’s official website to download the list of SpO

2

 probes and extension cords that 

can be used in conjunction with this device.

 Warning   

•  Check if the cable of SpO

2

 probe is in normal condition before measuring. After unplugging the SpO

2

 probe cable from 

the socket, “SpO

2

%” and “bmp” on the screen will disappear.

•  Do not use the SpO

2

 probe once the package or the probe is found damaged. Instead, you shall return it to the vendor.

•  The supplied SpO

2

 probe is only suitable for use with this device. This device can only use the SpO

2

 probe described 

in this manual. It is the responsibility of the operator to check the compatibility of the device and the SpO

2

 probe (and 

extension cable) before use. Incompatible accessories may result in device performance decreasing or cause injury to 

the patient.

•  SpO

2

 probe is a medical product that can be used repeatedly.

•  The measured value may be normal seemingly for the testee who has anemia or dysfunctional hemoglobin (such as car-

boxyhaemoglobin (COHb), methaemoglobin (MetHb) and sulfhaemoglobin (SuHb)), but the testee may appear hypoxia, 

it is recommended to perform further assessment according the clinical situations and symptoms.

•  Pulse oxygen has only reference significance for anemia and toxic hypoxia, as some patients with severe anemia still 

show better pulse oxygen measurements.

•  Measurement accuracy can be affected by the interference of electrosurgical equipment.

•  Do not install the SpO

2

 probe on an extremity with arterial catheter or receiving intravenous injection.

•  Do not perform SpO

2

 measuring and NIBP measuring on same limb at one time, because obstruction of blood flow 

during NIBP measuring may adversely affect the reading of SpO

2

 value.

•  Excessive movement (active or passive) of the subject or severe activity can affect the measurement accuracy.

ENGLISH

Summary of Contents for CONTEC08A-VET

Page 1: ...Holding Corp GmbH Europe Eiffestrasse 80 20537 Hamburg Germany Importato da Imported by Import par Importado por Importado por Gima S p A Via Marconi 1 20060 Gessate MI Italy gima gimaitaly com expor...

Page 2: ...e product In case of modifications and software upgrades the information contained in this document is subject to change without notice The warning items Before using this product you should consider...

Page 3: ...mit Prompt Function 56 8 1 Physiological parameter over limit prompt 56 8 2 Technical parameter over limit prompt 57 Chapter 9 The Usage Method of Sphygmomanometer 57 9 1 Accurate Measurement Way 57 9...

Page 4: ...42 Chapter 16 Keys and Symbols 70 Chapter 17 Maintenance Cleaning and Keeping 72 Chapter 18 NIBP Specification 74 ENGLISH...

Page 5: ...ication No Warning Self diagnosis and treatment using measured results may be dangerous Follow the instructions of your physi cian Please hand measurement results to the doctor who knows animal health...

Page 6: ...emble repair and modify the device without permission Otherwise it cannot be accurately measured This device can not be used on mobile transport platforms Otherwise it may cause measurement error This...

Page 7: ...t use it Otherwise it may cause fire or electric shock Please do not plug or unplug the adapter on the socket with wet hands Otherwise it may cause electric shock or injury 1 2 Operation for Battery N...

Page 8: ...te and record number etc With 2 8 inch color LCD screen clear interface the function of data review is complete User can implement ON OFF manual measuring system setup parameters change and other oper...

Page 9: ...ul life for five years since the first electrified use After its normal life dispose the device and its accessories according to applicable local regulations about environmental Chapter 2 MAIN UNIT Th...

Page 10: ...imal USB Data Line Software CD User Manual Optional Accessories AC adapter Input voltage AC 100 V 240 V frequency 50 Hz 60 HZ Rated current AC 150 mA Output DC 6 0 V 0 2 V 1 0 A or DC5 0 V 0 2 V 1 0 A...

Page 11: ...ay be useful to clinicians who perform optical therapy D Error in weak filling condition SpO2 and pulse rate can be shown correctly when pulse filling ratio is 0 4 SpO2 error is 4 when measuring range...

Page 12: ...airway tube plug away install it on the new cuff Note When the product and accessories described in this manual are about to exceed the period of use they must be disposed according to relevant produc...

Page 13: ...EC standards such as IEC60950 Information technology equipment Safety and IEC60601 1 Medical electrical equipment Safety and all equip ment should be connected to in accordance with the requirement of...

Page 14: ...ies power will exhaust Replace with four new batteries the same sort at the same time Test while low power may cause data deviation and other problems Turn the unit off before replacing the batteries...

Page 15: ...ed medical grade power adapter Note When regulated power supply and batteries are both used at the same time the battery power will not be con sumed Switch regulated power supply and battery as power...

Page 16: ...ery user If the date and time are set correctly the date and time when measuring will be correct in the memory otherwise it may not be correct The results can be uploaded to PC via USB and processed w...

Page 17: ...he year reaches 2099 pressing the UP button will return to 2010 Chapter 7 ABOUT UNIT There are two units mmHg and kPa The default is mmHg Enter the SYSTEM SETUP submenu in SYSTEM MENU then select UNIT...

Page 18: ...lood pressure When the BP measurement result is higher than the high limit or lower than the low limit and the prompt is ON the physiological prompt will occur in PROMPT SETUP interface select SpO2 PR...

Page 19: ...ence Advice Try to measure your animal s blood pressure at the same time each day with the same limbs and the same pose for consistency Do not touch Electronic Sphygmomanometer cuff and windpipe durin...

Page 20: ...flow and cause injury to the patient Do not use the cuff on the side of the mastectomy When using the cuff to pressurize some of the body s functions may temporarily weaken Use the measurement medica...

Page 21: ...nstructions below Ensure that the cuff is completely deflated Apply the appropriate size cuff to the animal Ensure that the cuff is not wrapped too tightly around the limb Excessive tightness may caus...

Page 22: ...M SETUP submenu in SYSTEM MENU and complete switching in CUFF TYPE item Note When the animals is a newborn please select the small cuff mode and select the appropriate size of the cuff to measure othe...

Page 23: ...ment process error messages will appear on the screen you can obviate the malfunction and restart a measurement The minimum value of the patient s physiological signal is the minimum limit that the de...

Page 24: ...igure shows that there is no measurement result can be displayed 3 Press LIST button to display the data list interface 4 Press TREND button to display trend interface End to display the measurement v...

Page 25: ...is temperature value is measured by a temperature measuring device The device does not provide over limit prompt function so it is inapplicable for using in places where need such function The SpO2 pr...

Page 26: ...use with this device This device can only use the SpO2 probe described in this manual It is the responsibility of the operator to check the compatibility of the device and the SpO2 probe and extension...

Page 27: ...Do not stare at the luminescent component directly when the device is turned on infrared light is invisible even if for maintenance purpose or it may have bad influence to the eyes Some models of func...

Page 28: ...is operation brings no affection to other functions Note SpO2 display range 0 100 PR display range 30 bpm beats min 250 bpm beats min If the SpO2 works abnormally after connecting the SpO2 probe to th...

Page 29: ...indocyanine green or methylene blue skin pigmentation It is required to use SpO2 probe which is provided by our company contact with our sale department when necessary Chapter 13 INSTALLATION OF THE...

Page 30: ...s Loose cuff Cuff is not connected correctly Correctly connect cuff refer to Chapter 9 Air leakage Cuff plug fall off Make sure the cuff plug is securely inserted in the windpipe refer to Chapter 9 Ai...

Page 31: ...asurement button Switch on the power once again and restart a measurement Power off suddenly when inflating No use for a long time the power of batteries can be exhausted owing to the changed temperat...

Page 32: ...late Stop using the device and contact us Other phenomenon Switch on the power once again and restart an operation Replace the batteries If no please contact us ENGLISH Symbol Description Symbol Descr...

Page 33: ...ass II applied Serial number Socket for power adapter Product complies with European Directive Authorized representative in the European community Product code Lot number Manufacturer Date of manufact...

Page 34: ...and accessories in liquid If any damage or deterioration of the device and accessories is found please do not use it Maintenance Clean the device and accessories regularly It is recommended to clean...

Page 35: ...ature humidity or dusty Aged cuff may result in inaccurate measurement please replace the cuff periodically ac cording to the user manual To avoid device damage keep the device out the reach of childr...

Page 36: ...in Resolution Pressure 1mmHg 0 133 kPa Accuracy Static pressure 3 mmHg 0 4 kPa Error The BP Value of the device is equivalence with the measurement value of Stethoscopy The error meets the requests in...

Page 37: ...hout batteries Safety classification Class equipment power supplied by power adapter Internally powered equipment power supplied by batteries Type BF applied part Service life The service life of the...

Page 38: ......

Page 39: ......

Page 40: ......

Reviews: