Contacta Loop-FSM User Manual Download Page 5

 

Measuring Frequency Response 

 

 

 

Detailed below are two  processes that can be used to con-

firm the system properly reproduces all of the required frequencies 
equally in the bandwidth of 100 to 5KHz. A test signal generator 
such as the Contacta Loop-TSG will be required to perform these 
tests and both methods will yield very accurate results. 

 
 
 

 
Method I  -  Sine wave signal 
 

Send at a minimum the following sine wave signals through 

the hearing loop system without adjusting any of the audio or loop 
level controls: 100Hz, 250 Hz, 500 Hz, 1000 Hz, 2500 Hz, and 5000 
Hz. Using the Flat Spectrum mode record the RMS level generated 
by each of these. To meet the IEC specification the levels should not 
vary by any more than ±3dB.  

 
 
 
 
Method II  -  Pink noise signal 
 

Send a pink noise signal into the hearing loop system and 

select the Third Octave mode on the meter. It will initially start at a 
1000Hz center frequency.  
 
 
 
 

Pressing button “B” will cycle through the frequencies of 

100Hz, 200Hz, 500Hz, 1000Hz, 2500Hz and 5000Hz. 

 

IEC 60118-4 Notes and Requirements 
 

The standard as revised notes that within the  frequencies of 

100Hz to 5000Hz the hearing loop system should equally reproduce 
all signals.  At a minimum the systems is to be tested at 100Hz, 1KHz 
and 5KHz. 

 

Program RMS/Peak Levels 

 
 

 

 

The second mode is used to set up the  hearing loop signal 

level using an A-Weighted filter. It is very difficult to set the field 
strength to an average of –12dBA as it will constantly vary based on 
the program used to set up the system. We will use the PEAK read-

ing to confirm that  both our design and equipment meet the IEC 
specification. Pressing  button “B” resets the PEAK reading. 

 
 
 

 
How to test then adjust the  loop system level: 
 

Once the hearing loop system has been installed, play some 

bandwidth limited (100Hz to 5KHz) pink noise or 1KHz sine (use the 
driver manufacturers recommendation) though the loop system. 
Walk throughout the audience area holding your meter vertically. 
Note the readings and confirm that the A-RMS level does not vary by 
more than ± 3dBA. This lets you know that the perceived signal level 
in the hearing aids will be the same, no matter where the person is 
sitting. 
 

Next: using program audio similar to what is normally used in 

the facility, adjust the audio program level to a level just above 
“normal”. Now turn on the hearing loop system and adjust the drive 
level until a –3dBA to 0dBA PEAK level is obtained in the center of 
the loop. Reset the  PEAK level and confirm this level averages 0 
dBA across the  audience listening plane. 

IEC 60118-4 Notes and Requirements 
 

The revised standard states that the signal level across the 

loop area should not vary by more than ±3dB. Once that is confirmed 
the signal level based on the facility’s normal program should peak 
at 0 dB as referenced to  a 400mA/m. This will confirm there is ade-
quate loop current to produce both the peak and average (Avg. is      
-12dB or 100ma/m) signal levels that work well with T-coil equipped 
hearing aids. 

Summary of Contents for Loop-FSM

Page 1: ...Model LOOP FSM User Manual Audio Frequency Induction Loop Field Strength Meter ...

Page 2: ...ion 0 1dB resolution for levels over 32dB Displayed resolution 0 1 dB Detection Type True RMS on all features Sensor Pickup Coil Direction of Sensitivity Length wise of meter s longer dimension and Parallel to meter face Position noted on Product Calibration Calibrated at 1 000 Hz sine to read 0 dB at 5 03 mG Frequency Response Flat 1dB from 50 Hz to 10 000 Hz A Weighted Class 2 meter specified in...

Page 3: ...p your own adapter use the following information 1 8 in TRS connector XLR female connector Tip Pin 2 Ring Pin 3 Sleeve Pin 1 3 Meter Operation Basics Turning the meter ON and Off On Press and hold button A until the screen lights Off Press and continue to hold button A until the Powering Down process is finished and the screen goes dark Selecting the meter mode When you press Button A it first dis...

Page 4: ...k noise signal Cont Record the RMS level readings for each of the frequencies As in method I if the level does not vary by more than 3dB the sys tem as installed will meet the IEC specification This method was requested by the field engineer so they could run the test with one instrument without continually adjusting the frequency source It also makes it easy to conduct the test in more than one l...

Page 5: ...o set the field strength to an average of 12dBA as it will constantly vary based on the program used to set up the system We will use the PEAK read ing to confirm that both our design and equipment meet the IEC specification Pressing button B resets the PEAK reading How to test then adjust the loop system level Once the hearing loop system has been installed play some bandwidth limited 100Hz to 5K...

Reviews: