background image

 

 

Set-up 

1.  Select [SETTINGS] by pressing 

MENU (F2)

2.  Follow the instructions on the screen to adjust language, contrast and automatic power-off timer or to 

erase the memory. Press 

OK or SELECT (F2)

 to confirm or 

CANCEL (F1) 

to exit. 

Measurement

 

1.  Connect the electrodes to the corresponding BNC inputs. Automatic temperature compensation and 

temperature measurements are possible with a Pt1000 temperature probe.  

2.  Press MENU (F2), select the desired mode with 



 and press 

OK or SELECT (F2)

 

Never immerse a pH electrode and a dissolved oxygen electrode together in the same solution! 

 

A blinking decimal point warns you for unstable measurements. Wait to read the display! Stirring the solution 
during the measurements promotes the homogeneity and is obligatory! 

 

Calibration 

Press 

CALIBRATE (F1)

 to go to the calibration menu 

 
The instrument will refuse automatic calibration when the electrode is unstable. Insufficient stirring or a worn 
electrode may be the cause. 

 
pH 

1.  The instrument permits to choose between the buffers in memory (1.68, 2.00, 4.00, 4.01, 6.87, 7.00, 

9.18, 9.21, 9.95, 10.01, 12.00, 12.45). Select the proper values and press 

OK.

 

2.  Rinse the electrodes with distilled water and immerse them in the first buffer solution. Select 

[CALIBRATE], press 

OK or or SELECT (F2) 

and follow the instructions on the screen until the 

calibration is finished. 

3.  After rinsing the electrodes with distilled water, immerse them in the samples and read the display. 
4.  Rinse the electrodes always with distilled water after use and store them in a 3...4 M KCl solution. 

mV  

1.  Immerse the electrodes in a standard solution of known potential. Select [CALIBRATE] and press 

OK or 

SELECT (F2)

. Adjust to the proper value and press 

OK or SELECT (F2)

 

Select [RESET] and press OK or SELECT (F2) to reset the calibration. 

Temperature 

1.  Immerse the Pt1000 in a solution of known temperature. Adjust to the proper value and press 

OK

 

Select [RESET] and press OK to reset the calibration. 

 

Without Pt1000, press OK, adjust the manual temperature compensation and proceed by pressing OK

Conductivity 

1.  The instrument permits to choose between the standards in memory (0.01, 0.1, 1 M KCl). Select the 

proper values and press 

OK or SELECT (F2)

 

2.  After rinsing the electrode several times with the first standard solution, immerse it in that standard. The 

solution temperature is not so critical but should lie between 0°C and 40°C. When no Pt1000 is used, do 
not forget to compensate manually first! Select [CALIBRATE], press 

OK or SELECT (F2) 

and follow the 

instructions on the screen until the calibration is finished. 

3.  Rinse the electrode several times with the sample, immerse it in that solution and read the display. 
4.  Rinse the electrode always after use and store it in distilled water. 

Dissolved oxygen  

1.  The display shows the air pressure correction and the salinity correction. Select the proper values and 

press 

OK or SELECT (F2)

. Leave salinity correction to zero unless you are going to measure in heavily 

salted solutions such as e.g. sea-water (35 g/l). Select [CALIBRATE], press 

OK or SELECT (F2) 

and 

follow the instructions on the screen until the calibration is finished. 

Summary of Contents for LabArt C51

Page 1: ...MANUAL HANDLEIDING MODE D EMPLOI ANLEITUNG Rev 03 2021 LabArt C51...

Page 2: ...C51 Input CH1 Conductivity TDS SAL Input CH2 pH mV Dissolved Oxygen Input C Temperature Pt1000 STORE F2 F1 OK...

Page 3: ...k 6 Data opslag in het intern geheugen 6 Information g n rales 7 Configure 8 Mesure 8 Calibration 8 Bonnes Pratiques de Laboratoire 9 Collection de donn es dans la m moire interne 9 Allgemeine Informa...

Page 4: ...hout charge provided the equipment has been used under normal laboratory conditions and in accordance with the operating limitations and maintenance procedures in this instruction manual and when not...

Page 5: ...rodes always with distilled water after use and store them in a 3 4 M KCl solution mV 1 Immerse the electrodes in a standard solution of known potential Select CALIBRATE and press OK or SELECT F2 Adju...

Page 6: ...on rate must be at least 10 cm s 4 Rinse the electrode always after use and store it in distilled water Good Laboratory Practice 1 Press CALIBRATE F1 to go to the calibration menu 2 Select GLP and pre...

Page 7: ...eden volgens de werkvoorschriften en onderhoudsprocedures uit deze gebruiksaanwijzing en wanneer de oorzaak van het defect niet een ongeluk een aanpassing verkeerd gebruik of misbruik is Er moet eerst...

Page 8: ...af 4 Na gebruik de elektroden steeds met gedistilleerd water reinigen en vervolgens in een 3 4 M KCl oplossing bewaren mV 1 Dompel de elektroden in een ijkoplossing van gekend potentiaal Kies IJKEN d...

Page 9: ...ktrode steeds schoon na gebruik en bewaar ze dan in gedistilleerd water Goede Laboratorium Praktijk 1 Druk op KALIBREREN F1 om naar het kalibratiemenu te gaan 2 Kies GLP en druk op OK om een volledig...

Page 10: ...s conditions normales de laboratoire selon les limitations op rationnelles et les proc d s d entretien de ce mode d emploi et que le d faut n est pas d un accident une adaptation un mal traitement ou...

Page 11: ...dans la solution mesurer Lisez la valeur donn e 4 Apr s usage rincez toujours les lectrodes avec de l eau distill e et immergez les dans une solution de 3 4 M KCl mV 1 Plongez les lectrodes dans une...

Page 12: ...lution mesurer Lisez la valeur donn e Il est obligatoire d agiter la solution pendant les mesures pour favoriser son homog n it La vitesse de courant doit tre au moins 10 cm s 5 Apr s usage rincez tou...

Page 13: ...malen Laboratoriumskonditionen und der Gebrauchsanleitung gem verwendet wurde nicht jedoch bei einem Unfall falscher Behandlung und Handhabung oder Modifikation des Ger tes Bevor ein Ger t w hrend der...

Page 14: ...lliertem Wasser werden diese in die Probenl sung eingetaucht und der Me wert kann abgelesen werden 4 Nach jedem Gebrauch die Elektroden mit destilliertem Wasser sp len und in einer 3 5 M KCl L sung au...

Page 15: ...Dabei sollte die Anstr mgeschwindigkeit mindestens 10 cm s betragen 3 Sp len Sie die Elektrode nach jedem Gebrauch und bewahren Sie sie in destilliertem Wasser auf Gute Laborpraxis 1 Dr cken Sie KALIB...

Page 16: ...erature Coefficient natural waters EN27888 SALINITY Range 0 70 ppt Resolution 0 1 ppt TDS Range 0 100 g l Resolution 0 1 mg l DISSOLVED OXYGEN Range 0 20 mg l 0 200 Resolution 0 05 mg l 0 5 Accuracy 1...

Page 17: ...0 9 21 10 01 12 00 5 C 1 85 3 99 7 08 9 45 10 24 12 56 10 C 1 89 3 99 7 06 9 38 10 18 12 35 20 C 1 96 3 99 7 02 9 26 10 06 12 14 25 C 2 00 4 00 7 00 9 21 10 01 12 00 30 C 2 03 4 01 6 99 9 16 9 87 11 8...

Page 18: ...DIRECTIVE 2004 108 EC EN61326 1 Consort bvba Parklaan 36 B 2300 Turnhout Belgium Tel 32 0 14 41 12 79 Fax 32 0 14 42 91 79 Sales sales consort be The electro magnetic susceptibility has been chosen at...

Page 19: ...epaid by sender provided the equipment has been used under normal laboratory conditions and in accordance with the operating limitations and maintenance procedures described in the instruction manual...

Reviews: