C931 • C932 • C933
44 / 47
Perform the following steps:
16.
reset the meter (switch on while holding MODE pressed).
yes?
15.
prepare the ion electrode (ISE) and solutions according to the manual.
yes?
18.
select the °C mode, measure temperature of solutions and read the display.
°C:
19.
connect the ISE, select the mV mode (do
not
calibrate!).
yes?
20.
rinse the ISE, dip in a fresh 1 ppm cal. so ISA (stir!) and read display.
mV:
17.
short-circuit the pH input, select the mV mode and read display.
mV:
21.
rinse the ISE, dip in a fresh 10 ppm cal. so ISA (stir!) and read display.
mV:
14.
type of connected electrode:
22.
rinse the ISE, dip in a fresh 100 ppm cal. so ISA (stir!) and read display.
mV:
23.
rinse the ISE, dip in a fresh 1000 ppm cal. so ISA (stir!) and read display.
mV:
Technical assistance form - Ion
If you need assistance with a technical problem, make sure to fill the
complete
checklist (all 23 points!) and fax a copy to Consort: +32/14/429179.
1.
purchase date:
5.
organisation:
7.
division:
6.
name:
8.
address:
9.
postal code:
10.
city:
11.
telephone:
12.
telefax:
13.
e-mail:
2.
indicate model:
4.
serial no:
C931 - C932 - C833
2.
purchased from:
"